a)Press the MOVE button until the arrow pointer is pointing at the delay menu item.
b)Press the CHANGE button. The delay submenu will appear. Rotate the ADJUST
knob until the delay is set at 1 us.
c)The arrow pointer should be pointing at the “Normal” choice. If it is not, press
MOVE until it is.
d)Press CHANGE to return to the main menu.
5.To set the pulse width to 1 us:
a)Press the MOVE button until the arrow pointer is pointing at the pulse width menu
item.
b)Press the CHANGE button. The pulse width submenu will appear. Rotate the
ADJUST knob until the pulse width is set at 1 us.
c)Press CHANGE to return to the main menu.
6.At this point, nothing should appear on the oscilloscope.
7.To enable the output:
a)Press the MOVE button until the arrow pointer is pointing at the output menu item.
b)Press the CHANGE button. The output submenu will appear.
c)Press MOVE until the arrow pointer is pointing at the “ON” choice.
d)Press CHANGE to return to the main menu.
8.To change the output amplitude:
a)Press the MOVE button until the arrow pointer is pointing at the amplitude menu
item.
b)Press the CHANGE button. The amplitude submenu will appear. Rotate the
ADJUST knob until the amplitude is set at 100V.
c)Observe the oscilloscope. You should see 1 us wide, 100V pulses.
d)Rotate the ADJUST knob. The amplitude as seen on the oscilloscope should vary.
e)Press CHANGE to return to the main menu.
9.Repeat the last step, but set the amplitude to zero.
10.This completes the first operational check test.
For the second test, the output module is connected.
28
Содержание AVO-6-B
Страница 38: ...PCB 158R5 LOW VOLTAGE POWER SUPPLY ...
Страница 39: ...PCB 197J HIGH VOLTAGE DISCHARGE BOARD ...
Страница 42: ...PCB 94H ALARM BOARD ...
Страница 44: ...PCB 104G KEYPAD DISPLAY BOARD ...
Страница 48: ...PERFORMANCE CHECK SHEET 48 ...