exceed 10 GHz.
d) Set the oscilloscope to trigger externally with the vertical setting at 50 mV/div
and the horizontal setting at 10 ns/div.
2. Turn on the AVMP-4-B. The main menu will appear on the LCD.
3. To set the AVMP-4-B to trigger from the internal clock at a PRF of 10 kHz:
a) The arrow pointer should be pointing at the frequency menu item. If it is not,
press the MOVE button until it is.
b) Press the CHANGE button. The frequency submenu will appear. Rotate the
ADJUST knob until the frequency is set at 10 kHz.
c) The arrow pointer should be pointing at the “Internal” choice. If it is not, press
MOVE until it is.
d) Press CHANGE to return to the main menu.
4. To set the delay to 100 ns:
a) Press the MOVE button until the arrow pointer is pointing at the delay menu
item.
b) Press the CHANGE button. The delay submenu will appear. Rotate the
ADJUST knob until the delay is set at 100 ns.
c) The arrow pointer should be pointing at the “Normal” choice. If it is not, press
MOVE until it is.
d) Press CHANGE to return to the main menu.
5. To set the pulse width to 60 ns:
a) Press the MOVE button until the arrow pointer is pointing at the pulse width
menu item.
b) Press the CHANGE button. The pulse width submenu will appear. Rotate the
ADJUST knob until the pulse width is set at 60 ns.
c) The arrow pointer should be pointing at the “Normal” choice. If it is not, press
MOVE until it is.
d) Press CHANGE to return to the main menu.
6. At this point, nothing should appear on the oscilloscope.
25
Содержание AVMP-4-B
Страница 27: ...This completes the operational check 27...
Страница 35: ...PCB 158R4 LOW VOLTAGE POWER SUPPLY...
Страница 39: ...PCB 151D RANGE POLARITY RELAY BOARD...
Страница 42: ...PERFORMANCE CHECKSHEET...