SPECIFICATIONS
Model:
AVH-HV1-B
2
Amplitude
3,9
: (50Ω load)
10 to 100 V
Pulse width, measured at
20% rise time
11
:
≤ 1.0 ns std.
(0.8 ns opt.
4
)
PRF:
0 to 100 kHz
Polarity
5
:
Positive, negative, or dual-polarity (specify)
Propagation delay:
(Ext trig in to pulse out)
≤ 200 ns
12
Required load impedance:
50 Ohms
10
Jitter:
± 15 ps (Ext trig in to pulse out)
Two channel option:
Optional
8
.
Double pulse option:
Optional
9
.
0 to ±5 ns pulse separation (other ranges available).
DC offset option
6
:
Apply required DC offset to back-panel solder terminals (± 50 Volts, 250 mA max)
Trigger modes:
Internal trigger, external trigger (TTL level pulse, > 10 ns, 1 kΩ input impedance),
front-panel “Single Pulse” pushbutton, or single pulse trigger via computer command.
Variable delay:
0 to 1.0 seconds, for all trigger modes (including external trigger).
Sync output:
> +3 Volts, > 50 ns, will drive 50 Ohm loads
Gate input:
Synchronous or asynchronous, active high or low, switchable.
Suppresses triggering when active.
Connectors:
Out: SMA, Trig, Sync, Gate: BNC
GPIB, RS-232 control
2
:
Standard on -B units.
Ethernet port, for remote
control using VXI-11.3,
ssh, telnet, & web:
Optional
7
. Recommended as a modern alternative to GPIB / RS-232.
http://www.avtechpulse.com/options/vxi
for details.
Power requirements:
100 - 240 Volts, 50 - 60 Hz
Dimensions (H x W x D):
100 mm x 430 mm x 375 mm (3.9” x 17” x 14.8”)
Optional rack-mount kit:
Add the suffix “-R5” to the model number to include 19” rack mount kit.
Operating temperature:
+5°C to +40°C
1) -C suffix indicates stand-alone lab instrument with internal clock and line powering. No suffix indicates miniature module requiring DC power and external
trigger. (See http://www.avtechpulse.com/formats for the basic instrument formats).
2) -B suffix indicates IEEE-488.2 GPIB and RS-232 control of amplitude, PRF, delay and polarity. (See http://www.avtechpulse.com/gpib).
3) For operation of variable-amplitude units at amplitudes of less than 20% of full-scale, best results will be obtained by setting the amplitude near full-scale and
using external attenuators on the output.
4) For 800 ps pulse width option, add suffix -T1.
5) Indicate desired polarity by suffixing model number with -P or -N (i.e. positive or negative) or -PN for dual polarity option. Polarity reversal achieved by means
of a two-position switch that controls the polarity of the signal output port on -C units and via keypad control on -B units. -PN option not available on modules.
6) For DC offset option suffix the model number with -OS. Avtech Model AVX-T bias tee can also be used to obtain DC offset.
7) Add the suffix -VXI to the model number to specify the Ethernet port..
8) For the two channel option, add the suffix -2CHPP for two positive outputs, the suffix -2CHNN for two negative outputs, or the suffix -2CHPN for the one positive
output and one negative output.
9) For the double pulse option add the suffix -DPP for a unipolar output, and add the suffix -DPN for a bipolar output. Note that the maximum amplitude is reduced
to 70% for DP option units.
10) A 50 Ohm load is required. Other loads may damage the instrument. Consult Avtech ([email protected]) if you need to drive other load impedances.
11) The FWHM pulse width, measured at 50% rise, will be lower.
12) Add 200 ns to propagation delay specification for units with the -DPP or -DPN double-pulse options.
6
Содержание AVH-HV1
Страница 36: ...PCB 235B HIGH VOLTAGE POWER SUPPLY...