GENERAL INFORMATION
BASIC TEST ARRANGEMENT
The AVB1-3 should be tested with a sampling oscilloscope with a bandwidth of at least
2 GHz to properly observe the high-speed waveform. A typical test arrangement is
shown below:
The AVB1-3 generates a positive pulse and a negative pulse internally. These are
added together in the pulse combiner located on the top of the main chassis. The OUT
P to P IN and OUT N to N IN cables shown in the diagram above are preinstalled on the
AVB1-3 (and should never be removed).
The attenuators are required to prevent damage to the sampling oscilloscope. A 40 dB
attenuator with sufficient voltage rating should be used on the main output.
6
AVTECH
AVB1-3
OUT N
SAMPLING
OSCILLOSCOPE
BW > 5 GHz
50 OHM INPUT
TRIG
TTL-LEVEL
LOGIC
PULSER
TRIG
MAIN OUTPUT
SYNC OUTPUT
40 dB
ATTENUATOR
ALL CABLES: 50 OHM COAXIAL
+3V to +5V (TTL)
0V
PW ≥ 50 ns
PRF ≤ 100 kHz
+15V DC
POWER
SUPPLY
50Ω FEED-THROUGH
TERMINATOR
(OPTIONAL, BUT
RECOMMENDED)
OUT P
P IN
N IN
OUT
N/C
T
PW
50Ω TERMINATOR
(PASTERNACK PE6071
OR EQUIVALENT, INCLUDED)