Note:

 If the patient’s skin is covered with hair, sweat, or dirt, clean the area 

and wait 10 minutes before taking a measurement. 

 

 

Note:

 Ensure that the thermometer is held firmly during measurement and 

that the patient does not move until the measurement is complete. Patient 
movement can impact the measurement. 

 

 
4. Press and release the 

Measure

 button making sure not to obstruct your 

    view of the LED display or the buttons area. 

 

 
5. Slowly move the device towards the patient’s forehead. The unit will 

    beep continuously and a series of dash icons forming two circles 
    will rotate on the display. Once the correct distance is reached 
    (about 1 inch or 3 cm), an amber light will flash just below the 
    

Mode

 and 

Memory

 buttons. The unit will emit a short beep 

    (except in night mode), and the backlit display will illuminate 
    when the measurement is complete. 

 

 

Note:

 If the distance of the thermometer is too far from or too close to the 

patient’s forehead, the dash icons will continue to rotate and the unit will 
beep (except in night mode) UNTIL the correct distance is obtained. Once the 

Measure

 button is pressed and released, you will have 20 seconds to obtain 

a temperature reading. After 20 seconds the unit returns to standby mode. 
 
 

 
 
 
 

 

 
6. To repeat the process, return to step 2. 

 

 
7. The thermometer automatically turns off (with a single beep) after 30 

    seconds of inactivity. 

 

 
Note:

 The device cannot be turned off manually.

Note:

  The Human or Object symbols will be displayed to indicate the  

measurement mode that was in use for the displayed reading in memory. 

 
3.  Each press of the 

Memory

 button will display a previous measurement 

    (up to 30).

10

11

To delete readings from memory:

 

1.  Press the 

Measure

 button to power the unit on.   

 
2.  Press and hold the 

Memory

 Button for at least 3 seconds 

     to delete all the readings. 

 
3.  The LCD displays  "--" and the device beeps to confirm 

     all the memories are cleared. 

 

Note:

  Once 30 measurements are in memory each new measurement will 

overwrite the oldest. 

 

Note:

  All the readings (Human and Object) will be cleared when deleting 

readings from memory.   
 

6.Memory Fucntion

 

The unit retains the last 30 readings.   

 

To recall previous readings: 

 

1.  Press the 

Measure

 button to power the unit on. 

 
2.  Briefly press and release the 

Memory

 button then press again to display  

    the last measurement accompanied by the             symbol.

93-429-00 Small IFU rev 6.qxp_Layout 1  4/1/20  12:06 PM  Page 10

Содержание ADTEMP 429

Страница 1: ... REP Distributed by American Diagnostic Corporation 55 Commerce Drive Hauppauge New York 11788 Inspected in the U S A Made in China tel 631 273 9600 1 800 232 2670 fax 631 273 9659 www adctoday com email info adctoday com 0598 93 429 00 Small IFU rev 6 qxp_Layout 1 4 1 20 12 06 PM Page 1 Contract Manufacturer AViTA WuJiang Co Ltd Tel 86 512 8207 8288 Address No 858 Jiao Tong Rd Wujiang Economic De...

Страница 2: ...ter is sure to provide you with many years of dependable service Please read the following instructions and general information which will prove helpful in allowing you to enjoy your ADC product Read this booklet thoroughly before attempting to use your new ADC Adtemp thermometer Thank you for your patronage It is indeed our pleasure to serve you Sincerely American Diagnostic Corp 2020 AMERICAN DI...

Страница 3: ...e for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people 14 Warranty This thermometer is guaranteed for a period of 24 months from date of purchase This warranty does not cover batteries or damage resulting from improper handling or use Specific legal righ...

Страница 4: ...PMENT and ME SYSTEM used will be connected to the PUBLIC MAINS NETWORK and the power input is otherwise within the scope of the Basic EMC standard Guidance and manufacturer s declaration Electromagnetic immunity Enclosure port a The equipment is suitable for use in Home Health Environments and Professional Health Care Environments limited to patient rooms and respiratory treatment facilities in ho...

Страница 5: ...els characteristic of a typical location in a typical commercial or hospital environment Guidance and Manufacturer s Declaration Electromagnetic emissions 16 5 2 Introduction Intended Use Intended use The Adtemp Non Contact thermometer is a clinical grade device intended for the intermittent measurement of human body temperature in patients of all ages It can also be used to measure the temperatur...

Страница 6: ...ution Do not modify or disassemble this device in any way without prior authorization Caution To avoid accuracy concerns make sure that the thermometer s lens is clean and intact prior to use and after measurement is complete Caution Avoid touching the lens directly with your fingers Caution Do not expose the thermometer to extreme temperatures or humidity levels Caution Do not expose to direct su...

Страница 7: ...thermometer for an excessive period of time or exposing the device to external heat sources may result in distorted temperature readings 14 7 10 Disposal The thermometer contains no hazardous materials Discard in accordance with national or local regulations Remove the batteries before disposal Battery Dispose of empty batteries in accordance with national or local regulations 9 Calibration Testin...

Страница 8: ...chieved using the following method CAUTION Never submerge the thermometer in water or any other liquid CAUTION Never use abrasive cleaning agents thinners or benzene for cleaning and never immerse the instrument in water or other cleaning liquids CAUTION Never insert a sharp object into the scanner area or any other open surface on the thermometer CAUTION Do not use unapproved cleaning or disinfec...

Страница 9: ...n 4c for instructions on selecting a different scan mode 3 Position the thermometer about 2 inches 5 cm from the center of the patient s forehead with the sensor aimed between the eyebrows 8 Maintenance 8a Replacing the Batteries The thermometer comes with two 1 5V AAA LR03 alkaline batteries Replace with two new AAA LR03 batteries when the low battery symbol appears on the display The thermometer...

Страница 10: ...on is pressed and released you will have 20 seconds to obtain a temperature reading After 20 seconds the unit returns to standby mode 6 To repeat the process return to step 2 7 The thermometer automatically turns off with a single beep after 30 seconds of inactivity Note The device cannot be turned off manually Note The Human or Object symbols will be displayed to indicate the measurement mode tha...