UltraSoundGate 116Unb          

 

User's Guide

 

part #31187 

RECORDER USGH Settings 

The configuration dialog box can be launched from the menu Options/
Configuration... or through the button. 

 

The input sample rate can be selected from the 

Input Device Set-

tings 

section. Available sample rates are 384, 256, 192, 128, 96 and 

48 kHz. The 

Buffer 

setting determines the USB transfer buffer size on 

the PC. Shorter durations will provide low real-time spectrogram dis-
play latencies but might lead to erroneous USB transfers under certain 
conditions. 

 

Generic USB Audio Device 

Getting started with the RECORDER USGH  
Software 

 

The RECORDER USGH software can be launched from 

Start / All 

Programs 

Avisoft Bioacoustics 

RECORDER USGH. 

On the 

first program start, the configuration dialog box w ill be launched 
automatically (otherwise it is available from 

Options 

Configura-

tion). 

Select the desired 

Sampling rate 

from the 

Input Device 

Settings 

section and click at 

Ok. 

Then click at the Pause button 

(

Monitoring/Pause

) and the Start button (

Monitoring/Start

). You 

w ill then see the real-time spectrogram displaying the incoming 
signals. For details on the operation of the RECORDER software 
see the Avisoft-RECORDER manual and the section 

RECORDER 

USGH Software Settings 

in this guide. 

Sound-activated recording can be arranged by selecting the 

Trig-

ger 

source option 

level of this channel

. The Pre-trigger duration 

should be kept as short as possible. Long pre-trigger settings can 
lead to data transmission errors if the sample rate is high. In order 
to simplify the operation of the Avisoft-RECORDER software in the 
field, a link to 

RECORDER USGH 

may be added to the Windows 

Startup folder 

(

Start->All Programs->Startup

). 

Additionally, the Avisoft-RECORDER option 

Monitoring/Autostart 

should be activated. This arrangement will start the monitoring 
process automatically after booting the laptop (the UltraSoundGate 
device must be attached to the USB port prior to booting Win-
dows). 

The menu 

Options/Configuration management/Presets 

provides 

a number of default settings that might  help to configure the sys-
tem for a specific task. Note that the relatively low gain of the Ultra-
SoundGate 116Unb requires that the trigger threshold is set to a 
low level. For passive bat monitoring try the preset 

Options / Con-

figuration management / Preset / Bat monitoring using 
USG116Hnm/Hnb/Hnbm

. It is also recommended to activate the 

option 

x10 

on the Display Settings section of the Configuration 

Dialog box that will provide a better (more sensitive) real-time 
spectrogram display. 

Please refer to the users guide or the online help system for further 
details on the Avisoft-RECORDER software (http://
www.avisoft.com/RECORDER.pdf). 