ZP900 Post Scales User Instructions

Introduction

Battery power

The optional battery pack connects to the two battery connections as shown in 

Figure 

2.8

. The battery provides approximately 20 hours of operation for a single display and 

15 hours for a dual display system before charging is necessary.

Figure 2.8  Battery pack connections

When a 24-36 VDC power supply is plugged into the power connection on the back 
panel of the ZP900, the green POWER light will turn on indicating power has been 
applied to the BSQ base. The red CHARGING light will blink to indicate the battery is 
charging. When the battery is fully charged the red CHARGING light will go out and the 
power supply can be unplugged to work on battery power only.

The standby power button, pointed out in 

Figure 2.8

can be used to turn the ZP900 

indicator on or off.

2.3  Powering up the ZP900

The indicator requires input power of 12 to 36 VDC @ 200ma min. The indicator is 
always ON as long as power is received. If using the ZP900 rechargeable battery pack, 
the battery timer setting can be used to turn the Indicator display OFF automatically.

Power can be supplied by:

l

The included AC to 24VDC in-line PSU connected to a properly grounded 
outlet (100 VAC - 240 VAC, 50 or 60 Hz)

l

Optional rechargeable battery pack mounted on the rear of the base. 20 
hours operation between charges. Recharge time is four hours using the in-
line PSU. The battery pack requires 24 to 36V to charge.

l

12 to 36 VDC power (150ma minimum at 24VDC) via a 2.1mm center 
positive barrel jack

Standby power 
button

Содержание ZP900

Страница 1: ...ZP900 Post Scales User Instructions AWT35 501429 Issue AF...

Страница 2: ...wo dimensional work and the making of a copy in two dimensions of a three dimensional work stored in any medium by electronic means or transmitted in any form or by any means including electronic mech...

Страница 3: ...ation 10 Removing the shipping stops 10 Leveling the scale 10 Connections 11 Connections on the ZP900 11 Connections on the BSQ 12 Powering up the ZP900 14 Front panel 15 Keys 16 Annunciators 16 Numer...

Страница 4: ...P AWTX supplied cable types 30 RS232 Serial ports default settings 31 For USA 31 USB HID 31 Ethernet 31 Chapter 6 Error messages 32 Chapter 7 Supervisor menu 33 Post application supervisor menu 33 Bat...

Страница 5: ...reflect the case of the key being described If a key has a dual function it may be referred to by its alternate function Displayed messages appear in bold italic type and reflect the case of the disp...

Страница 6: ...of protection provided by the socket is suitable for the environment N O U SER SER V I C EA B LE PA R TS R EFER TO Q U A LI FI ED SER V I C E PER SO N N EL FO R SER V I C E Equipment to be powered by...

Страница 7: ...user Take frequent breaks during prolonged usage 1 6 Sharp objects Do not use sharp objects such as screwdrivers or long fingernails to operate the keys IMPORTANT This equipment must be routinely chec...

Страница 8: ...se The type of indicators available include l base mounted l desk mounted l pole mounted display single or dual display available The indicator comes standard with two RS232 COM ports and can be upgra...

Страница 9: ...comes necessary in the future 2 1 2 Leveling the scale Place the scale on a level stable surface free from vibrations or strong air currents Level the scale using adjustable feet and the bubble level...

Страница 10: ...eyed RJ45 cable connector If substituting this with a non keyed RJ45 cable do not inadvertently connect between the BSQ Scale and ZP900 Ethernet connectors as it may damage the internal ZP900 Ethernet...

Страница 11: ...you choose One is for AC power only and the other is for use with the optional battery pack or AC power These two panels and their connections are shown in Figure 2 5 Figure 2 5 BSQ connections Power...

Страница 12: ...ly unit PSU is attached by a bracket to the back of the scale base The DC output barrel jack plugs into the 12 36 VDC input as shown in Figure 2 6 Figure 2 6 AC power block Plug the power cord into a...

Страница 13: ...wer supply can be unplugged to work on battery power only The standby power button pointed out in Figure 2 8 can be used to turn the ZP900 indicator on or off 2 3 Powering up the ZP900 The indicator r...

Страница 14: ...nd the display Figure 2 9 Indicator and remote display front panels F1 lb kg GROSS ZP900 ZERO GROSS lb kg Refer to the supplied label kit for appropriate capacity and classifier markings Labels should...

Страница 15: ...ey to access the password entry screen Menu Navigation Aborts a numeric entry and acts as an ESCAPE key for menu navigation Menu Navigation Move up in a menu Numeric entry Increments the value of the...

Страница 16: ...DOWN arrow key until 7 appears on the display Press the RIGHT arrow key again to shift the 17 left and insert a 0 Repeatedly press the UP arrow or DOWN arrow key until 9 appears on the display Press t...

Страница 17: ...hen in the configuration selection of a menu you must press either the ZERO key to accept or F1 to not accept the choice or value Then press the UP arrow key repeatedly until SAVE no is displayed 2 Us...

Страница 18: ...site location Use the Numeric entry procedure on page 17 The scale should then perform as though calibrated at the installation location Press F1 to bypass Gravity entry but this process will be repe...

Страница 19: ...s on while the indicator is in the menus and will change appearance according to the following rules All segments flashing This means you are in the menu structure but not in any of the following scre...

Страница 20: ...ghing mode press and hold the F1 key Enter password 111 and press the ZERO key 4 3 User menu The User menu is shown in Figure 4 2 Figure 4 2 User menu Use this menu to set the time and date to enter a...

Страница 21: ...ter values for the time h x Hour m x Minute s x Seconds StYLE Choose the style of the time display Choices are 12hr 12 hour clock 12hr AP 12 hour clock with AM PM 24hr 24 hour military time 4 3 2 Date...

Страница 22: ...ear 4 3 3 Site ID Site ID SitE Use this to enter a Site ID The Site ID can be used in a print format Valid entries are decimal 32 through 126 ASCII space to the character Enter Site ID 4 3 4 Seal Seal...

Страница 23: ...xxxxx Dload Sserial Dserial AWT30 XXXXX X X X XX AWT30 XXXXX X X X XX AWT30 XXXXX X X X XX Cell Sw Part Version Serial USB ETH MAC 1 xx 2 xx 3 xx 4 xx 5 xx 6 xx Activ8 USB ENET xxxxx xxxx xxxxx Refere...

Страница 24: ...Use this to view the version of the bootloader View bootloader PN View version 4 4 2 Firmware Firm Version Partno AWT30 XXXXX X X X XX PArtno Use this to view the firmware part number The part number...

Страница 25: ...Number of the indicator The number is displayed in two parts Press RIGHT arrow key or LEFT arrow key to toggle the display between the first and second parts of the serial number View SN 4 4 5 Enet E...

Страница 26: ...cell SN cELL Use this menu item to view information about the Quartz Digital Transducer QDT SW PArt View the software part number for the cell VErSion View the version of the cell software SEriAL Vie...

Страница 27: ...or Ethernet ports on the ZP900 Record the MAC address found in this menu before contacting AWTX for activation code instructions MAC View the MAC address of the indicator Ethernet port ACtiv8 Enter t...

Страница 28: ...t a change in the SYSTEM or PORTS menu has occurred 4 5 2 Scale1 cALib Display the number of times that a change in the CALIB menu has occurred conFig Display the number of times that a change in the...

Страница 29: ...e the USB or Ethernet ports 5 1 Serial communication port defaults Refer to the ZP900 Service Manual for full details on communication protocols and port configuration 5 1 1 Serial cable wiring for 9P...

Страница 30: ...7 to 6 and status bytes from 3 to 2 For certain software interfaces this may be unnecessary so test the communication before changing the default settings Refer to the ZP900 Service manual for details...

Страница 31: ...assword entry failed Z error Power up zero error see Power up zero on page 19 Gravity Power up gravity prompt Power up gravity on page 19 Error 250 Scale not communicating check interface cable from s...

Страница 32: ...ale and keypad cause battery power to be shutoff Values between 1 and 3600 minutes are valid Repeatedly press the UP arrow key until the indicator returns to normal weighing mode When battery timeout...

Страница 33: ...8 99 228 mm 11 99 305 mm 12 24 311 mm 1 29 33 mm MAX 55 14mm MIN 82 21 mm 25 6 mm Figure 8 1 Model 0912 BSQ dimensions with PSU but no indicator Figure 8 2 Model 0912 BSQ dimensions with PSU and indic...

Страница 34: ...m 13 74 349 mm 4 93 125 mm 08 78 mm 13 50 343 mm 14 50 368 mm 11 50 292 mm 1 61 41 mm FOOT 1 29 33mm MAX 55 14mm MIN 10 00 254 mm 12 00 305 mm 12 51 318 mm 82 21 mm 5 86 149mm MAX 5 37 136mm MIN 14 49...

Страница 35: ...line dimensions and connections Figure 8 5 Desk mount indicator dimensions BOTTOM VIEW 9 28 236 mm 4 43 113 mm 3 77 96 mm 8 43 214 mm DESKTOP SURFACE 4 46 113 mm 3 98 101 mm 22 5 5 mm 2X THRU ACCEPTS...

Страница 36: ...NG MOUNTING HOLES INDICATOR SHOWN 17 38 442 mm 3 29 84 mm 7 92 201 mm 4 72 120 mm 2 73 69 mm 12 83 326 mm 4 49 114 mm 49 12 5 mm 2 91 74 0 mm 27 7 mm 3X THRU ACCEPTS 6mm OR 1 4 SCREW FOR MOUNTING TO S...

Страница 37: ...TB3 pin 1 GND and pin 6 5V on the ZP900 indicator In order to connect to a PC terminal using a serial RS232 interface when two remote displays are already connected to the ZP900 serial ports you can u...

Страница 38: ...el bubble is located See placement in Figure 9 1 The NTEP label is placed on the right side of the base See placement in Figure 9 1 When certified a verification mark sticker certifying body supplied...

Страница 39: ...40 ZP900 Post Scales User Instructions Agency and sealing label positions...

Страница 40: ......

Страница 41: ...A Tel 507 238 4461 Fax 507 238 4195 Email usinfo awtxglobal com www averyweigh tronix com Avery Weigh Tronix UK Foundry Lane Smethwick West Midlands England B66 2LP Tel 44 0 8453 66 77 88 Fax 44 0 121...

Отзывы: