Malfunction of sensor or 
control system

GENERAL FAULT

(FLASHING)

ALARM ICON

INDICATES

ACTION

CAUSE

BLOCKED TUBE

(FLASHING)

WATER OUT

(FLASHING)

DISPOSABLE WATER 

PATH 

(FLASHING)

BATTERY CHARGING 

(FLASHING)

BATTERY 

(FLASHING)

GENERAL FAULT

(FLASHING) 

% O

2

DISPLAYS DASHES ( - - )

Cannot be corrected by user: 
disconnect patient. Shut off unit, 
send for service.

Internal component 
failure

O

2

 sensor fault

Reset by shutting off unit. Replace 
O

2

 sensor. Restart unit

Depleted or defective 
O

2

 sensor

High back pressure

Clear obstruction, check cannula 
type, re-install DPC

Obstructed or kinked 
cannula/delivery tube, 
incorrect cannula 
for flow rate or DPC 
improperly seated

No water in disposable 
water path. Gas flow 
continues without heating 
or water circulation

Disconnect patient. Replace water 
bag or straighten inlet tube. Restart 
unit.

Sterile water empty, or 
obstructed inlet tube

Disposable water path 
faulty or not detected. 
Unit will not run.

If disposable water path is present, 
remove and replace to reset 
detector.

Disposable water path 
defective, not properly 
seated or not installed

The internal battery backup is not fully charged. The unit would not run on battery for the full 
rated time in the event of a power failure. No action is necessary.

The unit is running 
BATTERY mode. Gas flow 
and blending continues 
without heat or water 
circulation.

Reconnect AC power

AC power is 
disconnected

High Flow (P04108)

5-40 liters per minute (lpm)

VAPCA, VAPCPAS, VAPCPS*

Low Flow (P04108LF)

1-8 liters per minute (lpm)

VAPCPM, VAPCN, VAPCI, VAPCII, VAPCISP, 
VAPCPS*

CANNULA FLOW RATES

CARTRIDGE

OPERATIONAL FLOW RATES

CANNULA TYPE

*VAPCPS is intended to deliver flow rates of 1-20 lpm