dose rate ranges:
range no.
1
2
3
4
5
6
7
8
analog scale start
analog scale centre
analog scale end
10
100
1000
0.1
1
10
1
10
100
10
100
1000
0.1
1
10
1
10
100
10
100
1000
0.1
1
10
unit
nSv/h µSv/h µSv/h µSv/h mSv/h mSv/h mSv/h
Sv/h
digital display format:
6150AD1/3/5 internal tube
_0.0_ -
99.9_
_20.0 -
_99.9
and
100_ -
999_
0.200 -
0.999
and
1.00_ -
9.99_
_2.00 -
_9.99
and
10.0_ -
99.9_
_20.0 -
_99.9
and
100_ -
999_
6150AD2/4/6 internal tube,
probe 6150AD-18
0.00_ -
9.99_
_2.00 -
_9.99
and
10.0_ -
99.9_
_20.0 -
_99.9
and
100_ -
999_
0.200 -
0.999
and
1.00_ -
9.99_
probe 6150AD-15
0.00_ -
9.99_
_2.00 -
_9.99
and
10.0_ -
99.9_
_20.0 -
_99.9
and
100_ -
999_
0.200 -
0.999
and
1.00_ -
9.99_
probe 6150AD-t
0.00_ -
9.99_
_2.00 -
_9.99
and
10.0_ -
99.9_
_20.0 -
_99.9
and
100_ -
999_
0.200 -
0.999
and
1.00_ -
9.99_
_2.00 -
_9.99
and
10.0_ -
99.9_
_20.0 -
_99.9
and
100_ -
999_
0.200 -
0.999
and
1.00_ -
9.99_
probe 6150AD-b
__0_ -
999_
0.200 -
0.999
and
1.00_ -
9.99_
_2.00 -
_9.99
and
10.0_ -
99.9_
When exceeding the highest range, the flashing full scale value (»9...9«) will indicate overrange.
Remark on the Teletector Probe 6150AD-t: ranges no. 2 to 5 are assigned to the low range tube, and
ranges no. 6 to 8 are assigned to the high range tube. When switching between these ranges, the 6150AD
switches between the two tubes accordingly.