36
Step 5:
T
ap on ʻʻCONFIRMʼʼ to save the alarm limits.
NOTE:
The minimum alarm limit cannot exceed the maximum alarm limit, even if the
maximum alarm limit is turned off. For example, if the maximum SpO
2
limit is turned
off but was previously set at 90%, the minimum SpO
2
limit cannot be set higher than
90%. If you want to set minimum SpO
2
limit at 90%, turn on the maximum SpO
2
limit,
set it above 90% and turn it off again as you wish.
NOTE:
The following table describes the default settings, adjustment ranges, and
intervals.
High Priority Alarm
Factory Default
Adjustment
Options
Adjustment
Interval
SpO
2
Upper Alarm Limit
Off
Off, 85-100%
1%
SpO
2
Lower Alarm Limit
85%
Off, 50-95%
1%
Pulse Rate
Upper Alarm Limit
Adult/
Pediatric
150 bpm
Off, 75-275 bpm
1 bpm
Infant
200 bpm
Pulse Rate
Lower Alarm Limit
Adult/
Pediatric
50 bpm
Off, 30-110 bpm
1 bpm
Infant
75 bpm
Temperature
Upper Alarm Limit
Adult/
Pediatric
98.6°F (37°C)
Off, 89.6-107.6°F
(32 to 42°C)
0.1°F
(0.1°C)
Infant
99.3
°
F (37.4
°
C)
Temperature
Lower Alarm Limit
Off
Off, 89.6-107.6°F
(32 to 42°C)
0.1°F
(0.1°C)
CAUTION!!!
The new ALARM LIMITS do not go into effect until the “CONFIRM” button is
tapped.
NOTE:
The upper and lower alarm limits appear as small-sized numbers next to the
measured value.