User and Technical Manual: HotDog Temperature Management Controller
Page 11
ALARMS AND ALERTS
NOTE: WARMING BLANKET OR MATTRESS WILL STOP WARMING IF AN ALARM CONDITION APPEARS
Error Codes
Error Description
Problem Solving Steps
E1
Over-Temperature
When the temperature exceeds one degree above the set point, the
Alarm sounds and power is removed from the Warming Device.
Unplug the Device to reset the Alarm. Wait 5 minutes and then
reconnect the Device. Turn the Controller on and select the
temperature. If the Alarm occurs again, stop using the Device and
contact technical support.
E2
Failure To Reach Temp
When the Warming Device does not reach the temperature set point
within 10 minutes, the Alarm sounds, and power is removed from the
Device. Check to make sure the Device is in contact with a patient
and the sensor area is touching the patient. Unplug the Device and
reconnect to reset. If the Alarm occurs again, stop using the Device
and contact technical support.
E3
Port Current Reached
If the electrical current in the Warming Device exceeds the allowable
limit, the Alarm sounds and power is removed from the Device. This
may indicate and electrical problem with the Device. Unplug the
Device from the Controller and reconnect to reset. If the Alarm
occurs again, stop using the Device and contact technical support.
E4
Sensor or Cable Failure
If the Controller loses communication with the sensor in the Warming
Device, an Alarm sounds, and power is removed from the Device.
This may be caused by an electrical problem in the Device or
Controller. Swap the cables and Device with known good product to
isolate the problem if possible. If the problem continues, stop using
the Device or cable and contact technical support
E5
Device Fold Detection
In Warming Device equipped with an over-temperature array, local
overheating caused by folding of the Blanket causes an alarm, and
power to the Blanket is turned off. Check the Blanket for folded areas.
To reset Alarm, unplug the cable, wait 5 minutes and reconnect. If
Alarm reoccurs, stop using the Blanket and contact technical support
E7
Auto Mode Disengaged
Ensure proper placement of the Temperature Sensor to continue use of
AUTO mode or manually control Warming Devices.
E8
Over-Temperature Sensor
(Secondary)
The temperature sensor has exceeded 46° C. Disconnect the Warming
Devices and contact technical support.
EA, EC, EF,
EH or EP
Hardware Failure
Please turn off the Controller, wait one minute, and then restart. If the
problem persists, contact technical support.
Contact Technical Support @ 952.465.3500 or [email protected]
Automatic Shutoff Timer - This feature prevents the System from unintentionally being left on.
The System is not
intended for use without a clinician present
. The System defaults to a 6-hour automatic shutoff, if there is no
interaction with the Controller. This time period can be adjusted for up to 24 hours, should it be required by the