2. Safety Information
2.6. Electromagnetic Immunity Declaration
Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the XVS System
Rated Maximum
Output Power of
Transmitter W
Separation Distance According to Frequency of Transmitter (m)
150 kHz to 80 MHz
outside ISM bands
150 kHz to 80 MHz in
ISM bands
80 MHz to 800 MHz
800 MHz to 2,5 GHz
0.01
0.12
0.2
0.4
1
0.1
0.37
0.64
1.3
2.6
1
1.17
2
4
8
10
3.7
6.4
13
26
100
11.7
20
40
80
Recommended Separation Distances (Headset and Optional Tablet)
Test
Frequency
(MHz)
Band
a
(MHz)
Service
a
Modulation
b
Maximum
Power (W)
Distance
(m)
Immunity
Test Level
(V/m)
Compliance
Level (V/m)
385
380 –390
TETRA 400
Pulse
modulation
b
18
Hz
1.8
0.3
27
27
450
430 – 470
GMRS 460, FRS 460
FM
c
± 5 kHz
deviation 1 kHz
sine
2
0.3
28
28
710
704 – 787
LTE Band 13, 17
Pulse
modulation
b
217
Hz
0.2
0.3
9
9
745
780
810
800 – 960
GSM 800/900, TETRA
800, iDEN 820,
CDMA 850, LTE Band
5
Pulse
modulation
b
18
Hz
2
0.3
28
28
870
930
1720
1 700 – 1 990
GSM 1800; CDMA
1900; GSM 1900;
DECT; LTE Band 1, 3,
4, 25; UMTS
Pulse
modulation
b
217
Hz
2
0.3
28
28
1845
1970
2450
2 400 – 2 570
Bluetooth, WLAN,
802.11 b/g/n, RFID
2450, LTE Band 7
Pulse
modulation
b
217
Hz
2
0.3
28
28
Test Specifications for Enclosure Port Immunity to RF Wireless Communications Equipment
(Headset and Optional Tablet)
2-9