The instrument only performs Tympanometry:
This is probably because of
the fact “Tymp” has been selected
in Power-
up under the Common Setup Menu instead of “Reflex
and Tymp”.
Can I review recorded reflexes?
Yes!
After a test session it is possible to review the recorded
reflexes. All recorded reflexes can be viewed in the small reflex
boxes in the lower part of the display. However, by pressing
“Man/Auto” (17) two small arrows “
” and “
” (F1 and F2)
makes it possible to browse between all recorded reflexes
from the small reflexes. When a specific reflex is viewed in the
small reflex box, the numerical data for the reflex test is
highlight and the graphical data of the reflex test is shown in a
large window above the reflex box.
What is Gradient?
Gradient is explained in the chapter “Tympanometry Setup
Menu” of this manual.
My AA222 only displays 4 reflexes boxes. I need 6 reflex boxes:
In the Common Setup Menu under Reflex Icon Boxes it is
possible to select between either 4 or 6 reflex boxes. Make
sure that you select the right number of reflex boxes according
to personal preferences.
I find the sensitivity of the reflex recording unsuitable! Can it be
modified?
Yes!
In the following Reflex Methods: Screening, Auto and
Sequence (Sequence only when Test Stop Criteria is set to
Automatic) it is possible to set the Reflex Sensitivity to
Sensitive, Normal or Robust.
In the chapter “Reflex Methods” in this manual, the three
settings are explained in detail.