P2/2

Buzzer on

Buzzer on

PV

PV

Figure 4. Absolute high alarm

Figure 5. Absolute low alarm

5.2 System Configuration Parameters 

 (accessed by code 0089)

Table 3. System configuration parameter setting

PSb

0

rd

1

CorF

0: 

° C  1:° F

     

End

Code

Setting Range

Initial 

Note

Inty

K

outy

3

inty

Atdu

Hy

                       

1

Description

outy

hy

atdu

psb

rd

corf

end

Input Sensor Type

Output Mode

Hysteresis Band

Ignore this setting

Input Offset

External alarm type

Display Unit

Exit

See Appendix 

0, 1, 2, 3, 4, 5, 6

0~9999
0~200

-1000~1000

0: low 1: high

1

Note 3

Note 4

Note 6
Note 7

Figure 6. The system setup flow chart

Note 3

. The value of outy should be 3. 

Note 4

. The default setting for the external buzzer is high limit alarm. The external 

b u z z er   w i l l   be  on  when  the  tempeature  is  above  SV+Hy  will  be  off  when  the 
temperature drops back to SV. The default setting for the Hy is 1.

Note 2. Set alarm for internal buzzer

The meter offers two alarms that can be set to turn on the internal buzzer at specific 
temperatures. The first alarm is controlled by parameters AH1 and AL1. The initial 
setting will turn on the buzzer at 481 ° F and off when temperature drops below 480° F. 
The second alarm is controlled by parameters AH2 and AL2. The initial setting of the 
second alarm is deactivated. It can be set as low alarm to send warning.
AH1 and AH2 are the temperatures to turn the inernal buzzer on; AL1 and AL2 are the 
temperatures to turn buzzer off. When AH1(2) >AL1(2), the alarm is set for absolute 
high alarm as shown in Figure 4 below. When AH1(2) <AL1(2), the alarm is set for 
absolute low alarm as shown in Figure 5 below. When AH1(2)=AL1(2), the alarm is 
deactivated.
Example, if AH1=481, AL1=480, when the temperature goes up to 481° F, the buzzer 
will be on; when the temperature drops down to 480° F, the buzzer will be off. If 
AH2=180, AL2=185, when the temperature drops down to 180° F, the buzzer will be 
on; when the temperature goes up to 185° F, the buzzer will be off.
User can press the shift key (>) to temporarily mute the buzzer sound. The alarm will 
buzz again if the alarm set temperature is reached again. To permanently deactivate 
the alarm, set AH1=AL1 or AH2=AL2. Please see flow chart in Figure 3 on how to set 
the value.

The default setting is both internal and external buzzer are on when the 
temperature is above 481 (SV+Hy, AH1); will be off when the temperature drops 
down to 480 (SV, AL1).

               

Operation Mode

0089

Enter Code

Parameter Display

Output mode selection

inty

X

    

XXXX

X

XXXX

outy

atdu

psb

Hysteresis Band

Input offset

SET

SET

SET

SET

SET

SET

SET

X

rd

Heating/Cooling

SET

X

corf

Display Unit(C/F)

end

SET

SET

SET

SET

SET

XXXX

HY

SET

Autotune offset

SET

Note 5

Note 5

. Calibration offset, PSb is used to set an input offset to compensate the error 

produced by the sensor. For example, if the meter displays 5 º C when probe is in 
ice/water mixture, setting PSb= -5, will make the controller display 0 ºC.

Note 6

. Rd is for system function selection, Its value should be kept as 1 for high limit 

alarm.

Note 7.

 Display unit selection CorF: 0 for Celsius (° C); 1 for Fahrenheit  (° F). 

Auber Instruments, Inc

5755 North Point Parkway, Suite 99

Alpharetta, GA 30022 USA

Tech support: 770-569-8420, 404-983-8228

Copyright 2007-2016, Auber Instruments All Rights Reserved.

No part of this manual shall be copied, reproduced, or transmitted in any way 

without the prior, written consent of Auber Instruments. Auber Instruments 

retains the exclusive rights to all information included in this document.

Additional note:
The remote alarm control output of this thermometer can also be used as a PID  or 
on/off control output. For details, please refer to Auber SYL-1614 manual. 