Instruction Manual

TH100ACHIMDX Thermometer 

for Stove Pipe, Chimney, w/ remote alarm

Version 1.1

  

2. Specifications

 

 AC adaptor: 100-240V, 50/60Hz input. 12 VDC, 1 Amp output.

 

 Power consumption: <2W

 

 Sampling rate: 4 samples/second

 

 Accuracy: 0.2% full scale

 

 Sensor: K type thermocouple, -320-1800° F (-200~900

°

C).

 

 Output for buzzer: 12VDC, 1 Amp

 

 LED display: 0.39 inch, red

 

 Internal buzzer: Two set points alarm.

 

 External buzzer: High limit or low limit alarm

 

 Dimension: 3x5x1.2"  (78x120x28.5mm)

3. Front Panel

8888

 

 AL - Internal Alarm Indicator

 

 Value increment / Select next parameter

 

 Value decrement / Select previous parameter

 

 Digit shift / Internal alarm mute

 

 Set / Confirm

 

 OUT - External buzzer indicator

 

 Parameter display

1. Features

This thermometer can be used for stove pipe and chimney temperature monitoring. 
The meter contains two alarm settings for the built-in buzzer. One can be used for high 
limit alarm  and the other can be used for low fuel alarm.  It also has a third alarm for 
the  connected  external  buzzer  that  can  be  placed  at  different  floor  from  where  the 
stove is located. The gauge is powered by 12V DC through an AC adapter (included) 
for continuous operation. It can also be powered by car battery. 

5. Parameter Setting

 

 POWER - 12V Power supply input

 

 

OUT - Output for external buzzer

 

 INPUT - Temperature probe input 

Figure 2 shows the terminals of the meter. Connect the 12V DC power adapter to the 
terminal 1 and wall outlet. Connect the external buzzer connector to terminal 2. The 
polarity for this socket is center pin positive (+), outer collar negative ( ). Connect the 
K  thermocouple  to  terminal  3.  Please  note  that  thermocouple  connector  also  has 
polarity. The wide blade should go to the wide slot.

4. Connecting the meter 

Note 1. Set Alarm for external buzzer.

There are two ways to set the external alarm  temperature.
a.  During  the  normal  operation  mode,  press  Λ  or  V  once  to  switch  the  display  from 
process value  (PV) to  set  value (SV,  or  target  temperature).  The  display  will  start  to 
blink.  Press  Λ  or  V  again  to  increase  or  decrease  the  SV.  When  finished,  wait  8 
seconds and the settings will take effect automatically (the display will stop blinking).
b.  Press  SET key  once.  Use >, Λ and V keys  to  enter code 0001.  Press  SET key  to 
confirm,  then  the display would be  SV (

Su

).  Press  SET key  again  to display the  SV 

setting.  Use >, Λ and V  keys  to  enter  the  new  SV  value  and press  SET  to  confirm. 
Press  V  key  to  change  the  display  to  END.  Then,  press  SET  to  exit.  You  can  also 
ignore the steps after confirmation of SV. The controller will return to normal operation 
mode automatically if no key is pressed for 1 minute. The flow chart below shows how 
to set the SV and alarms in details.

The default setting for the external buzzer is high limit alarm (see Note 6). The external 
buzzer will be on when the tempeature is above SV+Hy (see Note 5), will be off when 
the temperature drops back to SV. 

Symbol

Range

Note

SV External alarm temperature

                         

AH1

                            

Alarm 1 on temperature

End

Exit

AL1

                           

Alarm 1 off temperature

AH2 Alarm 2 on temperature
AL2 Alarm 2 off temperature

Description

Initial

SV

AH1

End

AL1
AH2
AL2

480
481
480
250
250

Note 1

Arbitrary value 

within the 

measuring range

Figure 3. Flow chart for how to set target temperature and alarm.

5.1. Temperature setting and Alarm setting (accessed by code 0001)

Table 1. Alarm parameters

Figure 1. Front panel

Figure 2. Terminals (back view)

Note 2

Operation Mode

XXXX

0001

Enter Code

Parameter Display

Buzzer On Temp

sv

    

480

0250

    

0250

External buzzer Tem

0249

ah1

al1

ah2

end

Buzzer Off Temp

Buzzer On Temp

SET

SET

SET

SET

SET

SET

SET

SET

    

0250

al2

SET

SET

SET

SET

SET

Buzzer Off Temp