User manual
– SYSTOE
14 168
Rev. G
Page 29/38
Battery pack SysToe
14 140
Double sided ring tape
14.3 Replacing the battery
Must be performed by a qualified service technician.
Tool required: screwdriver
To replace the battery, use a PZ head screwdriver to remove the four
battery compartment screws. Unplug the old battery.
Plug in the new battery and close the battery door. Tighten the screw firmly
but do not over tighten.
CAUTION
: The battery pack contains a temperature measurement
component that is matched to the cell type and to the product
requirements. Use only replacement battery packs as specified by Atys
medical.
CAUTION
: Unauthorized battery substitutions will void the warranty and
may cause overheating.
CAUTION
: Used NiMH batteries must be recycled or disposed of properly.