Instructions for Attaching the Device on the Body (ATP-C70)
►
State of the skin before attachment
An area larger than the device (ATP-C70) attachment area needs to be cleaned on the patient's skin.
If body hair is abundant in the attachment area, shaving will be required to remove the hair.
NOTE: If a wound or bleeding occurs when removing body hair, the device should be attached after
the bleeding stops.
As demonstrated in the "confirm patch
attachment site" figure
❶
, place the end of
the patch about a finger's distance below the
clavicle, and confirm the attachment site at a
45-degree angle.
If necessary, remove possible obstructions like
hair using a razor, etc. as shown in the "shave
the patch attachment site" figure
❷
.
Figure 1.3 Attachment preparation step
– Confirming attachment site
To produce the accurate ECG signals, clean the attachment area using alcohol swabs to remove any
possible obstructions such as dead skin cells, etc. from the skin surface. Then, fully dry the skin for at
least one minute before attaching the device (ATP-C70).
Cleansing the patch attachment site with alcohol
As illustrated in Figure
❸
, cleanse the attachment site to remove dead skin cells and possible
obstructions by wiping horizontally, vertically and diagonally with alcohol cottons or swabs. Then,
fully dry for at least one minute to make sure no alcohol is left on the surface.
Figure 1.4 Attachment preparation step
– Removing possible obstructions from the attachment site