ATS Bag Reinfusion Set Up And
Priming

A microemboli blood filter and I.V. blood set are required for

unwashed

blood reinfusion.

Caution: A new microemboli blood filter must be

used for each new ATS bag.

Priming of the blood filter and I.V. set is

accomplished by the following steps:

1. Prime I.V. blood administration set and microemboli blood filter with

sterile saline.

2. After chest drain disconnection, invert in-line ATS bag with spike port

pointing upward and remove tethered cap using sterile technique.
Insert saline filter spike into ATS bag spike port using a firm twisting
motion. Return ATS bag to upright position and place on standard
height I.V. pole.

3. Open filtered air vent located on top of ATS bag first, then open the

I.V. clamp to complete priming. All remaining air within the I.V. circuit
must be evacuated prior to patient connection. Close I.V. clamp when
fully primed. I.V. is now ready for patient connection. 

Caution: Fail-

ure to purge all air from the entire I.V. circuit prior to patient con-

nection can result in air emboli.

ATS Bag Reinfusion

Follow all hospital protocols for administering autologous whole blood
reinfusion for both gravity drip or pressure infuser application:

1. Attach distal end of fully primed I.V. set to patient and open I.V. line

clamp to begin patient infusion.

2. For non-pressure infusion, open filtered air vent for maximum flow

rate.

3. For pressure infuser application, filtered air vent must remain closed.

Maximum in-line bag infuser pressure is 150mmHg. 

Caution: Do

not reinfuse entire blood contents completely through blood filter

and I.V. set, as air emboli can result.

Caution: Anticoagulant therapy and dosage recommen   dations are

the discretion of a physician and should be monitored carefully dur-

ing and after patient reinfusion.

Have a question or need help in a hurry? 

Call Atrium toll free at 1-800-528-7486.

Visit us online atwww.atriummed.com

Reading And Recording Blood Volume

The in-line ATS bag deflects slightly under vacuum and is therefore
calibrated on one side under vacuum pressure. The opposite side is
calibrated for non-vacuum conditions. All fluid level calibrations are
in 25ml increments, up to a maximum collection volume of 600ml.
All ml increments are approximate.

Step 4: To Disconnect In-Line ATS Bag From Chest Drain 

To remove in-line ATS bag from the chest drain, securely close
patient tube clamp and both ATS bag clamps. Dis con nect “chest
drain” side first, then disconnect the “patient side” connector. Imme-
diately place the 

male 

patient tube con nector into the 

female 

chest

drain connector, and open patient tube clamp. Reconnect ATS bag
connectors to each other. ATS bag is now ready to be handled for
reinfusion use. 

Caution: Open patient tube slide clamp immediately following

recon nec tion to chest drain or new in-line ATS bag. Do not keep

patient tube clamp closed during chest drain age or patient transit.

Reconnect ATS bag
connectors to
each other

Open patient
tube clamp
after
reconnection
to chest drain

Keep
ATS bag
clamps
closed

Federal (USA) law restricts this device to sale by or on the order of a

physician. Refer to Instructions for Use for current indications, warnings,

contraindications, and precautions.

© Atrium Medical Corporation 2013.  All Rights Reserved  Printed in

U.S.A.  11/13 · Ocean  and Oasis are trademarks of Atrium Medical

Corporation, a MAQUET GETINGE GROUP company · 5 Wentworth

Drive, Hudson NH 03051 USA · 603-880-6718