The
cryo
ICE
BOX does not have any customer serviceable parts aside from mains fuses and
gas line desiccant filter for
cryo
ICE
BOX consoles so equipped. For servicing issues, contact
AtriCure, Inc. at:
www.atricure.com
7555 Innovation Way
Mason, Ohio 45040
USA
Customer Service/ Product Inquiries
Telephone: 513-755-4100
866-349-2342 Toll Free
Fax:
513-755-4567
Replacement of AC Line Fuses
Tools and Parts
• Needle Nose Pliers
• Fuses:
cryo
ICE
BOX
Model
Fuse Type
Manufacturer
Part Number
ACM1
T 4A L 250V
Schurter
0034.5049
ACM2
T 2A L 250V
Schurter
0034.5046
The
cryo
ICE
BOX unit has been pre-set at the factory to either 110V – 120V (ACM1) or
220V – 240V (ACM2). The Rating Label below the Power Entry Module on the back panel
of the
cryo
ICE
BOX indicates the selected Input Voltage for this unit. This setting should
only be adjusted by the manufacturer or by an authorized AtriCure technical service
representative.
NOTE:
cryo
ICE
BOX unit should be powered off and unplugged before continuing with
the fuse replacement procedure.
Procedure to replace AC Mains Fuses
1. Determine the fuse type by looking at the
cryo
ICE
BOX Model Number or the
cryo
ICE
BOX Rating Label.
2. Using the needle nose pliers, carefully extract the fuse box from the power entry
module by squeezing down on the fuse box tabs in the slots as shown in Figure 18.
xxvii