5.5 Network connection of the fusionTrack device
Connection of the fusionTrack to an IT-NETWORK that in-
cludes other equipment could result in previously unidenti-
fied RISKS to PATIENTS, OPERATORS or third parties. A
direct connection between the fusionTrack device and the
host PC must be used. Otherwise the integrator shall apply
chapter 14.13 of IEC 60601-1.
The fusionTrack uses a Gigabit Ethernet 1000BASE-T (IEEE 802.3ab) interface to communicate with
the PC. The fusionTrack factory setting uses 172.17.1.7 as the static IP address. The fusionTrack must be
properly powered (see Section 3.2.1) and connected to a PC.
The Ethernet adapter of the host PC must be configured as indicated in Table 5.4. For some hardwares,
not using jumbo frames may result in a much lower frame rate, see Table 5.5 for typical frame rates.
Address assignment
Static
IP address
172.17.1.100
a
Subnet mask
255.255.255.0
Jumbo frame
Enabled (at least 8500 bytes for maximal performance)
a
Actually, the last byte can be anything but 7, which corresponds to the device
IP.
Table 5.4: Default network settings.
Number of fidu-
cials
Gigabit
Ethernet
1000BASE-T,
1000 Mbit
/
s
100BASE-T, 100 Mbit
/
s
Ethernet
MTU
1500
bytes
(no
Jumbo Frame )
Ethernet
MTU
9014 bytes (Jumbo
Frame )
Ethernet
MTU
1500
bytes
(no
Jumbo Frame )
Ethernet
MTU
9014 bytes (Jumbo
Frame )
CPU
load
[%]
Network
[Mb
/
s]
Frame
Rate
[Hz]
CPU
load
[%]
Network
[Mb
/
s]
Frame
Rate
[Hz]
CPU
load
[%]
Network
[Mb
/
s]
Frame
Rate
[Hz]
CPU
load
[%]
Network
[Mb
/
s]
Frame
Rate
[Hz]
0
4
0.8
335
4
0.8
332
–
–
–
–
–
–
30
13
202
334
12
180
329
–
–
165
–
–
183
60
25
354
335
20
350
334
–
–
94
–
–
96
90
18
265
167
27
497
331
–
–
61
–
–
67
120
26
340
177
36
659
334
–
–
52
–
–
51
Table 5.5: Typical frame rates (Windows PC Intel i5-4690K). Green maximum performance, orange de-
graded performance.
The fusionTrack device can be used a Ethernet to USB adapter, however, the performances are not
guaranteed. Tests performed by Atracsys show that the Ethernet to USB adapters listing in Table 5.6 allow
a smooth usage of the GUI demo program.
5.6 Connecting and powering the fusionTrack
This section is an extract of the document ’fusionTrack electrical safety concept’. Upon request, a complete
copy can be provided.
Atracsys / 2020-06-17 / 16:48:00