Environmental parameter
Range
Temperature
−
10 °C to 55 °C
Atmospheric pressure
50 kPa to 106 kPa
Relative humidity
10 % to 90 % non-condensing
Table 5.3: Recommended long-term (1 week
−
6 months) storage and transportation conditions.
- keep away from strong heat or cold;
- keep safe from shocks, drops and vibrations;
- keep away from fluids;
- a ’fragile’ marking must be present on the package.
5.4 Mounting the fusionTrack device
The integrator has to perform tests related to instability haz-
ards §9.4 of IEC 60601-1.
The integrator of the fusionTrack system must take care of the mechanical stability of the fusionTrack at
all times (when the fusionTrack is in use but also when it is not powered). Effectively, the calibration of the
device may be invalidated if the device suffered from shocks, drop and strong vibrations. The fusionTrack
must therefore be firmly fixed to avoid any fall and should be mounted on a structure which minimises the
probability of shocks, drops and vibrations.
When installed on a tripod or on a cart, special care has to be taken to avoid tipping over. The shocks,
drops and vibrations have also to be avoided when the equipment on which the fusionTrack is installed
is moved like for example on a cart that passes through doors, is moved over a threshold or is put in the
parking position.
Firm mounting can be achieved by using either four (4) M4 screws (max depth 8 mm) or one (1) NC
1
4
′′
-20
screw. More explanations are provided on Figure 5.1
60
30
1/4- 20 UNC 7.95
28
M4 - 6H 8
Back
fi
xation plate
Figure 5.1: The fusionTrack screws fixation.
Atracsys / 2020-06-17 / 16:48:00
25 / 113