Matters need attention
A. Please read this instruction manual before using the detector.
B. The detector must be used in compliance with the designated procedures of our company.
C. The warranty period of this transmitter is 12months (starts from the date userget the detector).
Users should comply with the instructionsin the use of this detector. Any damages or malfunctions
caused by improperoperation of users are not within the scope of warranty.
D. Repairand replacement of any parts must be operated by trained sta
ff u
nder authorization of
our company, using original spare part. The operatoris responsible for the liability if users operate
the repair of replacement by themselves.
E. There is acid solution inside the transmitter, please do not disassembleit. And pay attention not
to damage the frontal membrane of the transmitter. If the skin is stained with the acid solution. the
right remedy is to flushing the stained skin continuously with clean water for 1 0minutes.
F. The detector must be calibrated for once each year. The explosion proof sheet parts of the
transmitter need to be cleaned regularly (blow with low pressure compressedair), otherwise the
detection sensitivity will be a
ff
ected by the dust and impurities clogging protection hole.
3
2
1.Power supply, signal line
2.Printed-circuit board
3.Terminal
4.Sound and tight alarm port
5.Air chamber of sensor
6.Sensor port
7.Main board connection
Ground
connection port
5
6
1
2
3
4
7
1. Key Technical Indicator
Repeatability: <2%
Zero drift: < ± 2% (F•S) /6M
Operating temperature: -20~+50
℃
Environment temperature: <90%RH
Working principle of the transmitter: electrochemical for oxygen and hazardous gases, catalytic
combustion for combustible gases
Lifespan of transmitter: 3years for combustible gas transmitter; 2years for hazardous gas
transmitter
2. Other Technical Parameters
Inspection mode: di
ff
usion
Working mode: fixed long-term continuous work
Working voltage: 12-30V DC
Working current: oxygen and hazardous gas <50mA (Max)
combustible gas <150mA (Max)
Technical performance and parameters
Notice to Reader
Please read this instruction manual before using the detector
Output signal: three-wire 4-20 mA standard signal output or RS485 signal output
Exterior structure: die-casting aluminum housing
Dimensions: 135 x 125 x 116mm (without mounting bracket)
167 x 145 x 126mm (with bracket)
Weight: without mounting brackets 1.2kg with brackets 1.6kg with stent
Connection cable (6 core cable) is defined as follows:
4-20mA with RS485:
Red: positive-input of power (12 to 24V)/ 24V+
Black: ground wire of power/24V
Yellow:4-20mA output.
Orange: RS485A
Blue: RS485B
Brown: Free
Connection distance: >1000m
Protection grade: IP65
Installation type: on wall
Type of explosion-proof: flame proof
Explosion-proof grade: Ex d IIC T6 Gb
Inlet port: hexagonal locknut
Settings of Remote Control
The remote control applied to all product type produced by Yuante.
1. Press "MENU" once to enter F01, address code setting menu. Press "OK" to modify, press"+""-
"to modify address code of the detector. Press "OK" to save settings and "BACK" to cancel.
2. Press "MENU", then press “+”to enter F02, minimum value setting menu of gas alarm, Press
"OK" to modify, press"+" "-"to modify the value. Press "OK" to save settings and "BACK" to
cancel.
3. Continue to press "+"to enter F03, maximum value setting menu of gas alarm., Press "OK" to
modify, press"+" "-"to modify the value. Press "OK" to save settings and "BACK" to cancel. Same
operation with F04 (zero point calibration), F05 (range calibration), F06 (AD value).
4. Press "MENU" four times to enter a shift status, under this status:
Press "RESET" once, the data would be 255, one more it would be 0. Just press "RESET" to shift.
When the data is 0, press "+"or"-"to modify alarm point, the data di
ff
erence would be only 1. When
the data is 255, press "+"or"-"to modify alarm point, the data di
ff
erence would be 1000. "RESET"
can also be used as mute button when it is alarming.
1
4
Fixed Gas Detector
YT-95H-:XX
Instruction Manual
Target gas
Range
Accuracy
least count
response time
O
2
0-30%VOL
<±2%(F.S)
0.01%VOL
≤
15
H
2
S
0-1000ppm
<±2%(F.S)
1ppm
≤
15
CO
0-1000ppm
<±2%(F.S)
1ppm
≤
25
H
2
0-20000ppm
<±2%(F.S)
1ppm
≤
15
CO
2
0-10000ppm
<±2%(F.S)
1ppm
≤
40
C
2
H
4
0-100ppm
<±2%(F.S)
0.01ppm
≤
60
NO
2
0-20.0ppm
<±2%(F.S)
0.1ppm
≤
25
N
2
0-100%VOL
<±2%(F.S)
0.01%VOL
≤
40
TVOC
0-6000ppm
<±2%(F.S)
1ppm
≤
60
NH
3
0-100ppm
<±2%(F.S)
0.01ppm
≤
40
ASH
3
0-10.00ppm
<±2%(F.S)
0.01ppm
≤
30
C
3
H
3
N
0-200.0ppm
<±2%(F.S)
0.1ppm
≤
40
F
2
0-10 ppm
<±2%(F.S)
0.01ppm
≤
25
O
3
0-5.00ppm
<±2%(F.S)
0.01ppm
≤
50
CL
2
0-10.00ppm
<±2%(F.S)
0.01ppm
≤
60
BF
3
0-10.00ppm
<±2%(F.S)
0.01ppm
≤
60
NC:Normally close
NO:Normally open
COM:Common
24V:24+,positive
GND:24-,negative
4~20mA:4~20mA analog signal
H
ig
h
A
la
rm
Lo
w
A
la
rm
R
el
ay
R
el
ay
Po
w
er
ed
Si
gn
al o
ut
pu
t
Motherboard connector
Alarm
Digital probe port
Digital probe port