![Atmos S 201 Thorax Скачать руководство пользователя страница 32](http://html.mh-extra.com/html/atmos/s-201-thorax/s-201-thorax_operating-instructions-manual_3003710032.webp)
32
4.6 Function
4.6.1 Target vacuum
•
Please note, an adjusted target vacuum over
-50 mbar may cause pain and injuries to the
patient.
• On the main screen the target vacuum can
be set directly by pressing the
,
or
55
Vac
buttons.
• ATTENTION: The change in the target
vacuum takes effect immediately. There is
no confirmation necessary.
•
The target vacuum can be freely selected
between -5 and -100 mbar in steps of 1
mbar.
•
If the buttons
and
are pressed
permanently, the increase / decrease is
accelerated
•
The target vacuum of -20 mbar is preset
when starting the device.
• The
55
Vac
button can be used to set the
target vacuum directly to -5 mbar.
For more information see “4.6.2
Schwerkraftdrainagemodus” on page 32
34.
If target vacuum is adjusted over -50 mbar the
notice appears ,,High target vacuum is set“.
4.6.2 Gravity drainage mode
•
On the main screen the target vacuum can be set directly to -5 mbar by pressing the
55
Vac
button.
•
If the
55
Vac
symbol is touched during inactive therapy, the target vacuum is set to -5 mbar. To
start the therapy, it must be started manually by pressing the symbol
.
•
If the
55
Vac
symbol is touched during active therapy, the target vacuum is set to -5 mbar.
NOTICE
)
The change of the target vacuum is immediately effective. There is no confirmation necessary.