5
User profile:
Doctors, medical auxiliaries without restrictions.
Patient groups:
Patients of all ages with and without restrictions.
Application organ:
Chronic and acute wounds on the skin which do not close by primary wound
healing.
Application time:
Under normal conditions for a continuous application over a period of more
than 30 days.
Application site:
Suitable for application in clinics, medical fields and home environment. The
application of the negative pressure device may only be conducted under physicians instructed
and by medically trained staff.
Contraindications:
Contraindications exist by patients with malignant tumour wounds (non
enteric fistulas) which have not yet been examined, osteomyelitis or necrotic tissue. The ATMOS
®
S 042 NPWT must not be applied to exposed nerves, anastomotic areas, blood vessels or organs.
The negative pressure device must not be used:
-
for non-medical applications.
-
in the presence of flammable or explosive gases or liquids.
The system is not evaluated for the emergency medicine in the rescue area (rescue vehicle, acci-
dent location).
The product is:
active
Sterility:
Secretion canister system is sterile.
Single-use product/reprocessing:
Secretion canisters are disposable products.
1.3 Function
The ATMOS
®
S 042 NPWT is a particularly convenient suction device for negative pressure wound
therapy. The vacuum unit is operated by an electromotive maintenance-free diaphragm pump.
During operation the pump creates a vacuum in the wound dressing, the hose system and the
secretion canister. This vacuum is used to suck secretions. The vacuum unit automatically cre-
ates, controls and monitors the set therapy values. The secretion is collected in the respectively
provided secretion canister. A special filter technology prevents penetration of liquids, bacteria
and odours into the device. Therapy settings can be made using the touch screen.
The vacuum unit is equipped with a rechargeable lithium-ionic battery. Safe recharging of the
battery is ensured by a microprocessor-controlled electronic charge system located in the vacu-
um unit.
The main performance feature is the regulation (in the normal case) and the limitation (in the 1st
fault) of the maximum possible negative pressure of the pump to approx. -500 mmHg.