3.5 Operating requirements
Please note that the following requirements must be adhered to for further operation after installing the device:
•
All joints and connection parts that are responsible for the safety of the device are securely fastened and fit
properly.
• All electronic connections (cables, plugs, power cables, etc.) are in good order and condition.
•
The line voltage and frequency specified on the microscope correspond to the values of the power supply.
•
The microscope is connected to a safety connection socket with the provided power cable.
Never point or direct the beam into the patient’s eyes. Do not look directly into the light source.
• With every light source, warming of tissue due to radiation and absorption could occur. This could result in damage to biological
tissue. Please keep the luminosity and duration of use to a minimum. Switch off the light source when not in use and check the
heat development if necessary.
!
3.0 Setting up and starting up
3.4 Starting up
•
Check whether the voltage values on the type plate correspond to the line voltage provided.
•
Check the scope of delivery.
•
Peruse the safety information in part 2.0 prior to starting up the device for the first time.
•
After transporting the microscope at low temperatures, it must be kept at room temperature for at least six hours before initial
start-up. If the microscope is
not
acclimatized, it
must
not be used.
•
Take into consideration when setting up the microscope that the elastic force of the arm without the microscope head is exceed
-
ingly strong.
Operate the brake of the height adjustment carefully.
•
To activate the ATMOS i View, please press the on/off switch on the front side of the control device.