Operating Instructions

English

503.0730.B

503.0700.B

2011-06  Index: 04

MedizinTechnik

Содержание Chair 21 P

Страница 1: ...izinTechnik GmbH Co KG Ludwig Kegel Str 16 79853 Lenzkirch Germany Tel 49 7653 6 89 0 Fax 49 7653 6 89 190 atmos atmosmed com www atmosmed com Operating Instructions English 503 0730 B 503 0700 B 2011 06 Index 04 MedizinTechnik ...

Страница 2: ... 7 8 3 1 Operating elements 7 3 2 Fixing the ATMOS Chair 21 P 7 3 3 Mounting the foot rest 8 3 4 Fixation of the back rest 8 3 5 Adjusting the head rest 8 3 6 Removing the head rest 8 4 0 Operation 9 4 1 Adjusting the seat height 9 4 2 Rotating the upper part 9 4 3 Adjusting the backrest 9 5 0 Cleaning and care 10 6 0 Maintenance and service 11 7 0 Troubleshooting 11 8 0 Options and spare parts 12...

Страница 3: ...t personnel authorised by ATMOS By applying only original spare parts you will have the guarantee that operational safety readiness for work and the value of your ATMOS Chair 21 P will be preserved The product ATMOS Chair 21 P bears CE marking CE according to the EC Directive of the council for medical products 93 42 EEC Directive 2007 47 EC dd 05 September 2007 and meets the basic requirements of...

Страница 4: ...d the position seating position or half lying position Application For treatment of human beings positioning Specification of the main function Height adjustment of the seat by mechanical hydraulics pump from approx 46 cm up to approx 66 cm 360 rotation Can be arrested in every height and every rotary position Backrest can be inclined from 5 up to 45 by gas com pression spring Height adjustable hea...

Страница 5: ...eration and battery charging 5 35 C 20 80 air humidity non condensing at air pressure 700 1060 hPa Graduierter Sekretbehälter 1l Sekretbehälter deckel mit 3facher Übersaugsicherheit REF SN 1 5 Transport and storage The transport of the device may be effected only in a dispatch carton upholstered and offering sufficient protec tion Please document and report damages in transit imme diately For compl...

Страница 6: ...es which are not approved by ATMOS for use with these chairs ATMOS neither guarantees error free function nor is ATMOS liable for personal injury and damage to property if no original ATMOS parts are being used the advice for use in these operating instructions is not being observed assembly new settings alterations extensions and repairs have been carried out by personnel not authorised by ATMOS ...

Страница 7: ... and start up Fig 2 Drillings for fastening of the base plate 3 2 Fixing the ATMOS Chair 21 P To guarantee safe operation of the ATMOS Chair 21 P it has to be screwed tightly on the floor This patient chair is fastened by means of the enclosed fixing set For this the cover plugs must be removed from the fastening drillings and the chair must be fixed to the floor with dowels and screws Fig 2 ...

Страница 8: ...ghtening them with the cylinder head screw M 8x20 and the spring washer Ø 8 1 Fig 4 3 5 Adjusting the head rest The height of the head rest can be adjusted by turning the screw Fig 5 It is possible to adjust the headrest in two different positions The headrest can completely be swivelled backwards and then be pulled out from its support Fig 6 Fig 5 Clamping screw Fig 3 Fixation of the foot rest 3 ...

Страница 9: ...to the first snap in position or even lower till the brake is completely released Then rotate the upper part of the chair with the patient and move the multifunction lever again in highest position rotation arresting when the desired position is achieved Fig 7 Multifunctional lever ATMOS Chair 21 P 4 1 Adjusting the seat height The swivel joint must be arrested multifunctional lever in uppermost po...

Страница 10: ...ng and disinfecting exterior surfaces Always observe the instructions for use by the manufacturer of the disinfectants above all concentration specifications and instructions relating to the material compatibility Covering material is made of synthetic leather Pay attention to material compatibility when cleaning or disinfecting the chair We recommend to clean the upper part of the chair after ever...

Страница 11: ...ns and repairs have been carried out by personnel not authorised by ATMOS The warranty will be rendered invalid in case of damages caused due to the utilization of accessories or consumables which are not approved by ATMOS for use with these chairs 6 0 Maintenance and service 7 0 Troubleshooting The ATMOS Chair 21 P was subjected to a close quality examination before leaving the factory If however...

Страница 12: ...inder head screw inner edge 6x30 051 0083 0 Cylinder head screw inner edge 6x20 051 0082 0 Washer 6 4 form b 053 0006 0 Base plate 503 0726 0 Counter sunk screw m 8x25 051 0387 0 Lifting cylinder 503 0716 0 Cylinder head screw inner edge 6x20 051 0082 0 Slide bearing iglidur g 503 0718 0 Slide bearing iglidur g 503 0717 0 Axis for backrest 503 0721 0 Axis for gas spring rear 503 0724 0 Axis for ga...

Страница 13: ... hPa Adjustment of the backrest 50 5 45 Rotation of the upper part 360 Dimensions HxWxD 115 x 58 x 64 cm Backrest HxW 58 x 54 cm Folding footrest HxW with optional folding foot rest 34 x 33 cm Head rest HxW 19 x 28 cm Diameter of the base 60 cm Weight 28 kg Max weight of the patient 150 kg Regular safety related inspections No Classification acc to Annex IX EC Directive 93 42 EEC Directive 2007 47 ...

Страница 14: ... be separated carefully Pay attention to country specific regulations for disposal e g waste incineration Disposal within the EC The procuct described above is a high quality medical product with a long service life After its life cycle it must be disposed of professional According to the EC directives WEEE and RoHS the device may not be disposed of in domestic waste Please observe existing nationa...

Страница 15: ......

Страница 16: ...sk of the goods damage or loss is therefore transferred to the client as soon as the goods leave the factory or the client is in default of acceptance of the goods This also applies to cases where we confirm prepaid carriage Transport packaging and all other packaging according to the packaging regulations is not returnable Our client is responsible for disposing the packaging at its own cost Our ...