- 49 -
Operating Manual Xenotest
®
220/220+
7
Operation
7.6 Settings for test programs
Table 4
ISO 105-B04
Irradiance (W/m
2
, 300 – 400 nm)
42 ± 2
Black standard temperature (°C)
Not higher than 20 Kelvin over CHT
Atlas recommends BST = 54 ± 2
for the programming
Test chamber temperature (°C)
CHT may not exceed 40 °C
during the drying period
Atlas recommends input 38 °C
for the programming
Duration of spraying (min.)
1
Duration of drying (min.)
29
Duration of weathering cycles (h)
16
Relative humidity
not controlled
Atlas recommends RH = OFF
for the programming