DC
High
Voltage
Probe
To
DMM
3
3. OPERATION
(1) Please connect ground lead and cables
probably before the measurements.
(2) Connect banana leads to DMM input.
(3) Set DMM to selected range (please use
fixed 10M ohm input impedance DMM).
(4) Caution!
Please do not turn on high voltage source
before completed above procedures.
(5) Turn on DMM and connect to the testing
point probably, and turn on high voltage
source.
4. WARNING
Do not attempt to take measurements from
sources when the chassis or return lead is not
ground.
This ground connection is critical to the safety
operation of the probe. Failure to make this
connection may result in personal injury or
damage to the probe or voltmeter. This
connection must be made before the probe tip
comes into contact with the high voltage and
must not be removed until after the probe tip
has been removed.
Do not connect the ground clip to the high
voltage source or the probe tip to the ground fo
any reason.
BEFORE turning the high voltage on, make
sure that no part of your body is in contact with
the device.
Remembering that the voltage being measured
is 1000 times greater than the voltmeter
reading.