Atlantic HVP-50 Скачать руководство пользователя страница 6

DC

 

High

 

Voltage

 

Probe

 

To

 

DMM

 

3

 

3. OPERATION 

(1) Please connect ground lead and cables 

probably before the measurements.   

(2) Connect banana leads to DMM input. 

(3) Set DMM to selected range (please use 

fixed 10M ohm input impedance DMM). 

(4)     Caution!  

Please do not turn on high voltage source 

before completed above procedures.   

(5) Turn on DMM and connect to the testing 

point probably, and turn on high voltage 
source. 

 

 

4. WARNING 

Do not attempt to take measurements from 
sources when the chassis or return lead is not 
ground. 

This ground connection is critical to the safety 
operation of the probe. Failure to make this 
connection may result in personal injury or 
damage to the probe or voltmeter. This 
connection must be made before the probe tip 
comes into contact with the high voltage and 
must not be removed until after the probe tip 
has been removed. 

Do not connect the ground clip to the high 
voltage source or the probe tip to the ground fo 
any reason. 

BEFORE turning the high voltage on, make 
sure that no part of your body is in contact with 
the device. 

Remembering that the voltage being measured 
is 1000 times greater than the voltmeter 
reading. 

 

www.tmatlantic.com