Model Q45CT Portable Transmitter
Part 4
– Configuration
23
O&M Manual
Rev-G (7/15)
TDS Factor
This function sets the linear relationship of the TDS (total
dissolved solids) reading to the conductivity measurement.
The actual units for the slope are in mg/L
/
μS. The default
value is 00.49 mg/L
/
μS.
Press ENTER to initiate user entry mode, and the entire
value will flash. Use the arrow keys to modify the desired
value; range is 00.00 mg/L
/
μS to 99.99 mg/L
/
μS. Press
ENTER to update and store the new value.
Out 2 Mode
This assigns the 4-20 mA output # 2 to Temperature (by
selecting 1),
S (by selecting 2) or mg/L (by selecting 3)
Temp Units
This function sets the display units for temperature
measurement. Press ENTER to initiate user entry mode,
and the entire value will flash. Use the UP arrow key to
modify the desired display value. The choices are
°F
and
°C
.PressENTER to store the new value.
4.10 Control Menu [CONTROL]
The Control Menu contains all of the output control user
settings:
Set 0V #1
Set 2.5V #1
Set 0V #2
Set 2.5V #2
These functions set the output range for each of the two
instrument outputs. The value stored for the 0V point may
be higher or lower than the value stored for the 2.5V point.
The entry values are limited to values within 0.00 and 14.00
pH and must be separated by at least 1% of this range. Use
the LEFT arrow key to select the first digit to be modified.
Then use the UP and LEFT arrow keys to select the desired
numerical value. Press ENTER to store the new value.
Output #1 will always be in units of pH, as it is fixed to track
pH. Output #2 will be in either units of pH or C/F, depending
on whether pH or temperature is set for Out#2 in the
CONFIG menu.