23
O&M Manual
Rev-C (05/20)
Part 4 – Operation
4.1
General
Check to see that sample is running through the flow assembly. Sample flow is controlled at 200
ml./min. If the system was not purchased with the flow control element, adjust the flow through
the system with a valve placed on the outlet side of the flow assembly. Recheck all wiring and
verify that the power supply is in the range of 12-24 VDC for external DC systems, and 12VDC for
battery systems. Note that battery powered systems will run from 12VDC down to about 7 VDC.
The controller does not contain a power switch. It will operate as soon as DC power is applied.
Once power is applied, the system will go through an initialization routine and then will begin to
operate.