Manual, F/T Sensor, Data Acquisition (DAQ) Systems
Document #9620-05-DAQ.indd-20
Pinnacle Park • 1031 Goodworth Drive • Apex, NC 27539 • Tel: 919.772.0115 • Fax: 919.772.8259 •
• Email:
27
Table 4.1—Unterminated Cable (9105-C-PS-U-x) for a NI 32 and 64 Pin Screw Terminal DAQ Device
NI 32 or 64
Pin # and Screw
Terminal Labels
Description
Wire Colors
NI 64 Pin # and
Screw Terminal
Labels
Description
Wire Colors
1
PFI 0/P0.0 (In)
No ATI Connection
N/A
33
PFI 8/P0.4 (In)
No ATI Connection
N/A
2
PFI 1/P0.1 (In)
No ATI Connection
N/A
34
PFI 9/P0.5 (In)
No ATI Connection
N/A
3
PFI 2/P0.2 (In)
No ATI Connection
N/A
35
PFI 10/P0.6 (In)
No ATI Connection
N/A
4
PFI 3/P0.3 (In)
No ATI Connection
N/A
36
PFI 11/P0.7 (In)
No ATI Connection
N/A
5
D GND
No ATI Connection
1
N/A
1
37
D GND
No ATI Connection
1
N/A
1
6
PFI 4/P1.0 (Out)
No ATI Connection
N/A
38
PFI 12/P1.4 (Out)
No ATI Connection
N/A
7
PFI 5/P1.1 (Out)
No ATI Connection
N/A
39
PFI 13/P1.5 (Out)
No ATI Connection
N/A
8
PFI 6/P1.2 (Out)
No ATI Connection
N/A
40
PFI 14/P1.6 (Out)
No ATI Connection
N/A
9
PFI 7/P1.3 (Out)
No ATI Connection
N/A
41
PFI 15/P1.7 (Out)
No ATI Connection
N/A
10
+5V
No ATI Connection
1
N/A
1
42
+5V
No ATI Connection
1
N/A
1
11
D GND
Shielding
Twisted shielding
43
D GND
Shielding
2
Twisted shielding
2
12
NC or AO 0
No ATI Connection
N/A
44
NC
No ATI Connection
N/A
13
NC or AO 1
No ATI Connection
N/A
45
NC
No ATI Connection
N/A
14
Reser or AO GND
No ATI Connection
N/A
46
AI GND
No ATI Connection
N/A
15
AI 0
SG0 output
Brown
47
AI 16
SG0 output
2
Brown
2
16
AI 8
SG0 reference
Brown/White
48
AI 24
SG0 reference
2
Brown/White
2
17
AI 1
SG1 output
Yellow
49
AI 17
SG1 output
2
Yellow
2
18
AI 9
SG1 reference
Yellow/White
50
AI 25
SG1 reference
2
Yellow/White
2
19
AI 2
SG2 output
Green
51
AI 18
SG2 output
2
Green
2
20
AI 10
SG2 reference
Green/White
52
AI 26
SG2 reference
2
Green/White
2
21
AI 3
SG3 output
Blue
53
AI 19
SG3 output
2
Blue
2
22
AI 11
SG3 reference
Blue/White
54
AI 27
SG3 reference
2
Blue/White
2
23
AI SENSE
No ATI Connection
N/A
55
AI GND
No ATI Connection
N/A
24
AI 4
SG4 output
Violet
56
AI 20
SG4 output
2
Violet
2
25
AI 12
SG4 reference
Violet/White
57
AI 28
SG4 reference
2
Violet/White
2
26
AI 5
SG5 output
Grey
58
AI 21
SG5 output
2
Grey
2
27
AI 13
SG5 reference
Grey/White
59
AI 29
SG5 reference
2
Grey/White
2
28
AI GND
AGnd power input
Black
60
AI GND
AI GND
2
Black
2
29
AI 6
No ATI Connection
N/A
61
AI 22
No ATI Connection
N/A
30
AI 14
No ATI Connection
N/A
62
AI 30
No ATI Connection
N/A
31
AI 7
No ATI Connection
N/A
63
AI 23
No ATI Connection
N/A
32
AI 15
No ATI Connection
N/A
64
AI 31
No ATI Connection
N/A
Notes: (Note # shown in table as superscript)
1. Do not connect the (+5V) red wire and the power ground (DGND) red/white wire on systems that have PS or
IFPS boxes with a 12 V wall mounted power supply. The (+5V) red wire and the power ground (DGND) red/white
wire are only required for systems that do not have a separate wall mounted power supply. Connect to external
power source, not the +5V through the USB.
2. Indicates the wired connection for the 2nd transducer and PS or IFPS box for NI 64 Pin screw terminal
connections only.