6
Electrical specification for the +/-15VDC P7A connector are in the following table:
Table 2.2—IF-OEM
±
15 V Electrical Specifications
Signal
Minimum
Typical
Maximum
Units
+ V
ANA
Power Input Voltage
13.0
15.0
17.0
V DC
- V
ANA
Power Input Voltage
-17.0
-15.0
-13.0
V DC
+V
ANA
Power Input Current
N/A
50.0
mA
-V
ANA
Power Input Current
-45.0
mA
V
ANA
Power Input Noise
75.0
mV p-p
V
ANA
Power Input Regulation
0.5%
N/A
NOTICE:
Custom cables that mate to the P7A and P7B headers use Molex microminature
1.25 mm receptacles.
2.3 P7B Connector for Signal Output
P7B is the output for the six amplified signals that connect to the DAQ device. The signal output is in the
following table:
NOTICE:
The T REF and T OUT signal is used only on custom sensors with a thermistor
installed. Standard sensors do not have a thermistor; do not connect these signals to
the DAQ device.
Table 2.3—P7B Signal Output
Pin Number
Color
Signal Name
Note
1
Brown/White
G0 REF
}Twisted Pair
2
Brown
G0 OUT
3
Yellow/White
G1 REF
}Twisted Pair
4
Yellow
G1 OUT
5
Green/White
G2 REF
}Twisted Pair
6
Green
G2 OUT
7
Blue/White
G3 REF
}Twisted Pair
8
Blue
G3 OUT
9
Violet/White
G4 REF
}Twisted Pair
10
Violet
G4 OUT
11
Grey/White
G5 REF
}Twisted Pair
12
Grey
G5 OUT
13
White/Black
T REF
}Twisted Pair
14
White
T OUT