14
Atmospheric Technology
stored data includes: start time; stop time; starting pressure; ending pressure;
average, minimum, and maximum flow rate; error status, and elapsed time. The
entire data set can be downloaded through the jump drive by inserting the drive in
the USB port and pressing the
Store
button. The file names on the jump drive are
unique and previously stored files do not have to be deleted before storing new data.
The data file name is the port or channel number and the date/time when the data
was written to the jump drive. An example of a typical data file is shown here:
East River Site
Canister 13249
Port 1 Canister Started at 11/20/00 9:00:06 PM
Starting Pressure 0.11 psia
Flow Rate Set Point 3.29 cc/min
Stopped at 11/21/00 9:00:05 PM
End Pressure 11.68 psia
Total Sample Time 24.00 hours
Average Flow Rate 3.29 cc/min
Minimum Flow Rate 0.94 cc/min
Maximum Flow Rate 3.47 cc/min
Flow Controller Zero -0.02 cc/min
Error Code 4
Error Status Pressure Tolerance Exceeded
Time Flowrate Pressure
11/20/00 9:00:17 PM 0.34
0.26
11/20/00 9:05:26 PM 3.28
0.16
11/21/00 8:40:15 PM 3.28
11.53
11/21/00 8:45:25 PM 3.29
11.61
11/21/00 8:50:25 PM 3.28
11.63
11/21/00 9:00:05 PM 3.27
11.68
Data for each port is saved until the port is sampled again. The new data is stored
in the location of the previous data. All data can be erased after a sampling
sequence by answering “Yes” to the query that appears when exiting the Data
screen.
The
Next
button on the 8001 Data screen will advance to the next data screen or
Exit
can be used to return to the main screen.