13
CAUTION
• Be sure to use sterilizers that can perform sterilization no
higher than 275°F (135°C). In some sterilizers, the
chamber temperature may exceed 275°F (135°C). Do not
use these sterilizers as failure of the motor could occur.
Contact the sterilizer manufacturer for detailed information
about cycle temperatures.
• Steam sterilization is recommended for the product. The
validity of other sterilization methods (such as plasma
sterilization or EOG sterilization) is not confirmed.
7-5
Storage
Store the product in a dry, clean location.
CAUTION
• After the sterilization and drying cycles are complete,
remove the motor immediately from the sterilizer to store it.
• Store the product in a well ventilated place out of direct
sunlight and within the range of temperature, humidity and
pressure specified in “10. Specifications”.
• Sterilization is not guaranteed after the sterilization
retention period specified by the manufacturer and seller of
the sterilization pouch has elapse.
If the sterilization retention period has elapsed, perform
sterilization again with a new sterilization pouch.