plenty of time (minimum 30 minutes) to stabilize to the working/room temperature.
‘Er’: Error 5~9, for all other error messages it is necessary to reset the thermometer. To reset, turn the instrument off, 

remove the battery and wait for a minimum of one minute, reinsert the battery and turn on. If the error message remains 

please contact the Service Department for further assistance.
BATTERIES

The thermometer incorporates visual low battery indication as follows:

‘Battery OK’: measurements are possible

‘Battery Low’: battery needs to be replaced, measurements are still possible

‘Battery Exhausted’: measurements are not possible

When the ‘Low Battery’ icon indicates the battery is low, the battery should be replaced immediately with AAA, 1.5V 

batteries. Please note: It is important to turn the instrument off before replacing the battery otherwise the thermometer may 

malfunction. Dispose of used battery promptly and keep away from children.
SPECIFICATIONS

Measurement Range

-60 to 550˚C (-76 to 1022˚F)

Operating Range

0 to 50˚C (32 to 122˚F)

Accuracy, (Tobj=15-35˚C,Tamb=25°C) ±1.5˚C (2.7˚F)
Accuracy, (Tamb=23 ±3˚C)

Tobj=0 to 550˚C: ±2% of reading or 2˚C (4˚F) whichever is greater 

Tobj=-60 to 0˚C: ±(2˚C+0.05/˚C)

Emissivity

0.95 default - adjustable 0.1 to 1 step .01

Resolution (-9.9 to 199.9˚C)

0.1˚C/0.1˚F, otherwise 1˚C/1˚F

Response Time (90%)

1 sec

Distance:Spot

12:1

Battery Life

Typ. 18, min 14 hours continuous use 

Dimensions

112.58 x  147.93 x 41.36mm (4.43x5.82x1.63”)

Weight

145 grams (5.1 oz) including batteries (AAA*2 pcs)

Note: Under the electromagnetic field of 3V/m from 200 to 600MHz, the maximum error is 10˚C (18˚F.)
EMC/RFI: Readings may be affected if the unit is operated within radio frequency electromagnetic field strength of approxi-

mately 3 volts per meter, but the performance of the instrument will not be permanently affected.

252224-INST-ATD