PiezoClamping
Calibration Kit
Page 2 of 14
www.atcp-ndt.com
Copyright
Copyright © 2020 by ATCP Physical Engineering
All rights reserved.
ATCP reserves the right to change the product and this manual without previous
notice.
Version 1
August / 2020
Страница 1: ...ions and Operation Manual PiezoClamping Calibration Kit ATCP Physical Engineering Rua L da Vassimon n 735 A Ribeir o Preto SP Brazil 14 026 567 Telephone 55 16 3289 9481 E mail ha atcp ndt com www atc...
Страница 2: ...tion Kit Page 2 of 14 www atcp ndt com Copyright Copyright 2020 by ATCP Physical Engineering All rights reserved ATCP reserves the right to change the product and this manual without previous notice V...
Страница 3: ...and symbology 5 3 Operation principle nominal charges and uncertainty 6 4 Parts accessories and optional items 7 5 Technical specifications and elements identification 8 6 Configuration and use 9 7 Ba...
Страница 4: ...g the equipment Its misuse may compromise the results and cause damage PiezoClamping Calibration Kit provides standardized electrical charges 5 50 11 00 16 50 and 22 00 C for testing the instrument Pi...
Страница 5: ...e in electric potential between two places It is what makes electric charges move The unit for electrical potential difference or voltage is the Volt V Capacitor The capacitor is a component which has...
Страница 6: ...options are 2 0 5 0 7 5 and 10 0 Volts This approach provides a progressive charge and voltage levels PiezoClamping Calibration Kit simplified schematics PiezoClamping Calibration Kit top and side vi...
Страница 7: ...accessories and optional items PiezoClamping Calibration Kit is provided with the items described next Parts PiezoClamping Calibration Kit A Accessories BNC BNC cable B USB cable C Specifications and...
Страница 8: ...ions W x L x H 70 x 62 x 50 mm Equipment weight without packaging 85 g Working temperature range From 10 to 45 C 5 2 Elements identification 1 ON OFF toggle switch Switch to turn the equipment ON and...
Страница 9: ...rocedure Step 01 Turn the PiezoClamping equipment on Then press or to access the charge display screen as shown in the figure below Step 02 Connect the Calibration Kit connector OUT 4 to the PiezoClam...
Страница 10: ...on PiezoClamping to tare the equipment and start a new charge measurement Step 06 Change the toggle switch OUT CHARGE to OUT in order to discharge the selected charge in the PiezoClamping Step 07 Read...
Страница 11: ...1 00 C 5 10 45 C 11 55 C 3 16 50 C 5 15 67 C 17 32 C 4 22 00 C 5 20 90 C 23 10 C As the values measured by PiezoClamping are within Table 2 acceptance ranges the PiezoClamping equipment is calibrated...
Страница 12: ...hese steps to charge the PiezoClamping Calibration Kit battery Step 01 Switch the Calibration Kit OFF using the toggle switch ON OFF 1 Step 02 Connect a smartphone charger not included to the USB char...
Страница 13: ...measurements are not reproducible The BNC BNC cable is not connected Connect the BNC BNC cable The BNC BNC cable presents poor contact or one of the internal wires has been interrupted Replace the BN...
Страница 14: ...may invalidate the warranty agreement Lack of care as recommended by this manual in what regards the installation and operation of the equipment Accident fall inadequate installation or any other dam...