Adjusting to the Reference
m
Before adjusting to a reference solution, confirm that the prism surface is clean.
①
Confirm that the sample inlet unit is
properly connected to the piping.
②
Let tap water or a reference sample flow
through the piping.
③
Supply power to the instrument referring
to the procedure described in Chapter 9
"Power Supply" on page24.
The current DMF(%) is displayed.
④
Press the
for one second.
[t] is displayed.
Press the
key again and the display
will change to [1] (Fig. 13-1).
⑤
When the DMF(%) value blinks, adjust the
value to 0.0% for tap water or to the true
value for the reference sample by using
the
and/or
keys (Fig. 13-2).
⑥
Pressing the
sets the adjustment
and the display returns to [1] (Fig. 13-1).
⑦
Each time the
is pressed, the menu
item displayed switches in the order of:
[2], [3], [4], [5], and DMF(%).
Select DMF(%).
N
If 30 seconds pass when in steps ④
through ⑥ above, the display
returns to the continuous DMF(%)
display.
To reset the reference point to original factory
settings
①
While the DMF(%) is blinking in step ⑤
above, press and hold down both the
and
keys at the same time.
After 5 seconds, the display will clear.
Release the keys and the original
standard DMF(%) will be displayed.
②
The reference point is now reset to
original factory settings.
Fig. 13-1
Example of the adjustment to the 0.0%
Fig. 13-2