24
Use with a data logger
BF5 User Manual v 2.3
2 Heater power available, BF5 permanently
enabled
This connection would be appropriate to an installation with
permanent mains power available. The BF5 heater is powered up,
and the BF5 analogue outputs are permanently enabled for logging.
Note
The key point is that the various grounds (Htr-, DL-Gnd, Signal Gnd)
are connected at the BF5, so if the heater is in use, then the Signal
Gnd must be able to float relative to the two power grounds at the
logger end.
This is achieved easily with a differential input.
For a single-ended logger, the requirement is that either the logger or
the heater PSU must be able to float relative to the other, and there
must be no connection between signal and power grounds or BF5
enclosure anywhere except inside the BF5.