background image

3

Viewing aperture

in center of

Cheshire eyepiece

Reflective surface

of Cheshire

eyepiece

Secondary mirror

spider vane

Secondary

mirror in its

holder

Reflection of the

interior of central

baffle tube; the

glare stop baffles

are visible as light

concentric rings

Figure 1:

View through a

Cheshire eyepiece

(not to scale)

Viewing aperture is

off center in reflective

surface of Cheshire

Figure 2:

Secondary mirror

out of collimation

(not to scale)

Secondary mirror is

centered when viewed

through Cheshire

End of focuser

drawtube (optical axis)

Surface of

room wall

Front end of

optical tube

Front end of

optical tube

End of focuser

drawtube (optical axis)

SECONDARY

MIRROR

COLLIMATION

SCREWS

Collimation screws

(4mm hex head)

DO NOT ADJUST

CENTER

PHILLIPS-HEAD SCREW

DO NOT

ADJUST

Matte black interior

of optical tube

telescope at a white (or light colored) wall. Remove all of the

extension rings and attach the focuser directly to the optical

tube. Insert the Cheshire eyepiece fully into the focuser using

the 1.25” eyepiece adapter. Lock the focuser drawtube firmly

in place. Make sure there is a light source directed at the 45°

cutout in the side of the Cheshire.

Look through the Cheshire eyepiece. You will see a small

black dot within a centrally-located bright circle as seen in

Figure 1, above. The central black dot is the viewing aperture

in the center of the Cheshire eyepiece. The bright circle around

the central dot is the 45° reflective surface of the Cheshire

eyepiece and the larger black circle surrounding that is a

reflection of the interior of the scope’s baffle tube in the

secondary mirror. Your room wall and the interior of the optical

tube form the background.

Concentric light-colored rings will be visible in the black

circle of the secondary mirror if your room and the light source

aimed at the cutout in the side of the Cheshire are bright

enough. These are the reflections from the glare stops

machined into the baffle tube interior. You are seeing the front

of the glare stops that face the sky and their visibility here

simply shows that they are doing their job of reflecting stray

light back at the sky.

The outermost ring of light around the entire Cheshire field,

as shown in Figure 1, is the end of the focuser drawtube (the

optical axis of the scope). You can disregard this for the time

being. It will be covered later, when checking the primary

mirror collimation.

If the central black dot (the viewing aperture of the eyepiece)

appears centered in the circular reflective surface of the

Cheshire eyepiece as shown above, no further significant

adjustment of the secondary mirror will be necessary.

Secondary Mirror Collimation:

 However, if the black dot

of the viewing aperture appears off-center as in Figure 2, at

the top of the next column, adjust the three secondary mirror

collimation screws shown below until the viewing aperture is

centered as closely as possible in the Cheshire’s circular

reflective surface.

A user-supplied 4mm hex key is required to collimate the

secondary mirror. Adjust only the three hex head screws around

the perimeter of the holder, as shown in the illustration in the

previous column. 

Do not adjust the center screw. Adjusting

the center screw can cause the secondary mirror to fall off

and any resulting damage will not be covered under warranty.

As you adjust each of these screws you will need to make

equal counter-adjustments to the other two. In other words,

as you tighten one screw you will need to loosen, by an equal

amount, the other two. The opposite is also true. If a screw is

loosened, the two opposing screws should be tightened. When

the process is complete you should have equal tension on all

three screws.

Only minor adjustments should be required to fine-tune

the collimation. Adjust the screws no more than an eighth of

a turn or less at a time. This will help prevent accidently putting

the optics grossly out of collimation. The force vector diagram

on the next page will show you how different adjustments

affect the tilt of the secondary mirror.

The correct alignment of the secondary mirror is critical in

determining if the optical axis (primary mirror) requires

alignment. Be certain you have properly aligned the secondary

mirror before proceeding to the next step of adjusting the

optical axis collimation, using the primary mirror collimation

screws shown below.

Optical Axis (Primary Mirror) Collimation: 

As

mentioned above, the optical axis of the scope

 

(the primary

mirror/baffle tube assembly) will rarely need collimation. If

the optical axis does get knocked out of collimation, however,

the image through the Cheshire eyepiece will appear to be

shifted to one side within the light ring formed by the end of

the focuser drawtube, as shown in Figure 3 on the next page.

If properly collimated, all of the light and dark circles will be

concentric, as shown in Figure 1 in the previous column.

Adjusting the optical axis will require user-supplied 3mm

and 2.5mm hex keys. There are three pairs of “push-pull”

hex-head screws on the rear cell of the optical tube, as shown

AT8RC OPTICAL AXIS

COLLIMATION

SCREWS

Collimation screw

(2.5mm black

hex head)

Lock screw (3mm

silver hex head)

Rear cell

Отзывы: