Page 26
Version 0.7
Additionally, MGBox can also send information about calibration values stored in the EEPROM.
Standard Firmware:
$PCAL,P,20,T,50,H,-10,MG,0,MM,0*73<cr><lf>
10Micron specific Firmware:
$PCAL,P,20,T,50,H,-10,UR,1,UT,0*7E<cr><lf>
$
Start of sentence
P
Proprietary sentence
CAL
Calibration data
P
Sensor type, pressure
20
Calibration value multiplied by 10, e.g. 2.0
hPa
T
Sensor type, temperature
50
Calibration value multiplied by 10, e.g. 5.0 °C
H
Sensory type, humidity
-10
Calibration value multiplied by 10, e.g.
-1.0 %RH
MG / UR
MG: Send GPS-Data to Mount
UR: Update mount’s refraction data
1
Enabled (0 = Disabled)
MM / UT
MM: Send Meteo-Data to Mount
UT: Synchronize mount’s time
0
Disabled (1 = Enabled)
*78
Checksum
Setting, resetting, receiving Calibration data and other commands
In order to set, reset or query calibration data from the MGBox, similar formatted sentences can
be sent to the device:
Setting pressure calibration
:calp,50*
:calp
Set calibration value for pressure
50
Calibration value multiplied by 10, e.g. 5.0 °C
*
End of message indicator
Setting temperature calibration
:calt,22*
:calt
Set calibration value for temperature
22
Calibration value multiplied by 10, e.g. 2.2
hPa
*
End of message indicator