Code 0000137992
Rev 1.0
2
INSTRUCTION SETTING
Functions
:
CALIBRATION
Press
for 3 Seconds
Sequence Standard calibration buffer solution Redox 465mV
SET POINT
Press:
hold down
and change the value with key
750mV
Press
(together) for 5 Seconds and run Program Setup
PROGRAM
Press
to set the following Item:
Language
With the button
It’s possi
ble to have 5 language EN, IT, ES, DE, FR
Flow
Select with the button
It’s possible to enable(ON) or disable (OFF) signal
input flow (high voltage)
connected in parallel to the recirculation pump.
SETPOINT 750mV
Select with the button
and set with
It’s possi
ble to adjust from 0 to 1000 mV for Redox.
SETPOINT TYPE HIGHT
Select with the button
and set with
It’s possible to adjust dosing and High or Low
OFA TIME OFF
Select with the button
and set with
It’s possible to adjust OFA time in minut
es
CALIBRATION 465mV
Select with the button
and set with
It’s possible to select
465 mV for the redox function or disabled.