Workmanship - AST, Inc. pressure transmitters have a limited one-year warranty to the original purchaser.  AST, Inc. will replace or repair, 

free of charge, any defective transmitter.  This warranty does not apply to any units that have been modified; misused, neglected or installed 

where the application exceeds published ratings. AST’s sensors are made with pride in New Jersey, USA. If in the area please feel free to 

stop by for a visit!

Installation/Applications

 - The purchaser is responsible for media compatibility, functional adequacy, and correct installation of the transmitter.

TM

AG200 Operating Instructions

Loop Voltage

Select a loop power supply voltage and total loop resistance so that when the loop 

current is 20 mA, the gauge will have at least 7.5 VDC at its terminals. Too large 

a loop resistance will cause the gauge output to “limit” or saturate before reaching 

its full 20 mA output.

The minimum loop supply voltage may be calculated from the formula:

Vmin = 7.5V + (20mA x Total loop resistance).

If the loop voltage drops too low, the AG200 will display the LOW LOOP icon.  When 

the LOW LOOP icon is displayed, all calibration functions, described below, are dis-

abled.

Normal Operation

Under normal operation the AG200 displays the pressure in the selected units and 

outputs 4mA to 20mA current proportional to the pressure.  The AG200 outputs 4mA for 

zero or low end pressure, and 20mA for full scale or high end pressure.  If the pressure 

reading goes outside the specified range, the AG200 will output as low as 3.5mA or as 

high as 24mA loop current.

Press the UNITS button the change the displayed units.  The AG200 displays PSI, 

bar, Atm, Kg/cm2, InHg, InH2O and mA loop current.  The UNITS button does not 

affect the current output.

Generating a Test Current

When the TEST button is held depressed, the display and loop current are switched to a 

test level.  This test mode will allow setup and testing of the current loop by switching to 

this test level whenever desired without having to alter the system pressure.  To change 

the test output level, press the CALIB button, while the TEST button is depressed.  

Each time the CALIB button is pressed, the test level current increases by 2mA.

Zero Pressure Calibration

The gauge port must be open with no pressure or vacuum applied.  Press the ZERO 

and CALIB buttons together for about 3 seconds to zero the unit.  The buttons can be 

released when the gauge shows a zero pressure reading.  Note:  It is best to perform 

this step when pressure units (not mA) are shown on the display.  The zero calibration 

is retained after the unit is turned off.

Span Pressure Calibration

Span calibration should only be attempted if the user has access to a pressure 

reference of known accuracy.  The calibration equipment should be at least four 

times the gauge accuracy.  Zero calibration must be done before span calibration.   

To perform span perform the following steps:

1. Choose the calibration units by using the UNITS button.

2. Hold the CALIB button for about 3 seconds.

3. The display will alternate between CAL and the required calibration pressure.  For 

instance, if the unit needs to be calibrated at 500 PSI, the display will show: 

CAL

 then 

500

   

4. Connect the unit to a pressure source and apply the pressure indicated on the dis-

play.

5. Press and hold the 

CALIB button for about 3 seconds to complete the calibration.  If 

the calibration is successful, the display will show: 

Done

If there is an error, the display will show: 

Err

An error condition will occur if the user tries to calibrate the unit /- 10% 

of the 

factory calibration.

• To cancel the calibration at any point during the procedure, press and release 

the 

CALIB button.

• Factory calibration can be restored by entering the calibration state, as described in 

step 2, then holding the ZERO button for about 3 seconds.  The unit will display:  

FAC

to indicate that factory calibration has been restored.  

Current Output Calibration

The AG200 has been set at the factory to be in agreement between the displayed value 

and the 4-20mA loop current. These settings should not normally be 

adjusted.  If adjustment is necessary, perform the following steps:

1. Choose the mA display units, using the UNITS button.

2. Connect the loop to an accurate current measurement device.

3. Hold the CALIB button for about 3 seconds.

4. The display will alternate between CAL and 4.00mA:

CAL

 then 

4.00

the loop current will be set to 4.00mA.

5. Use the 

TEST button to decrease the output loop current by 0.01mA.  Use the UNITS 

button to increase the output loop.  Note that even though the output loop current will 

change, the display will not.

6. To accept the calibration at 4.00mA, press and hold the CALIB button for about 3 

seconds, until the display changes, as described in step 7.  You can also cancel any 

changes you made, by pressing and releasing the CALIB button.

7. The display will now alternate between CAL and 20.00mA:

CAL

 then 

20.00

the loop current will be set to 20.00mA.

8. Again, use the TEST button to decrease the output loop current by 0.01mA.  Use the 

UNITS button to increase the output loop current by 0.01mA.  Note that even though the 

output loop current will change, the display will not.

9. To accept the calibration at 20.00mA, press and hold the CALIB button for about 3 

seconds, until the display shows: 

DONE

You can also cancel any calibration changes you made, by pressing and releasing the 

CALIB button.

Factory calibration of the loop current can be restored by entering the loop current 

calibration state, as described in step 3, then holding the ZERO 

button for about 3 seconds.  The unit will display:  

FAC

to indicate that factory calibration has been restored.