Setting up a test connection on the HYAMP III is straightforward, but differs
slightly by model.
3140/3145/3160 Set Up
3130 Set Up
Connect the black return test lead to the black Return terminal located on
the front panel of the instrument.
For 3140/3145/3160 models
connect the sense lead to the Sense(-)
terminal located just above the return terminal.
Clip the end of the black return test lead to the chassis ground of
the DUT.
Connect the red high current test lead to the red Current terminal on the
front panel of the instrument.
For 3140/3145/3160 models
connect the sense lead to the Sense(+)
terminal.
Connect the end of the red high current test lead clip to the ground/earth
contact of the DUT.
Push the provided current and return test leads into their mating terminal until
seated against the rubber grommet. Avoid excessive force in seating. If the
grommet becomes compressed, the lock feature may not activate. To release the
lock and remove the leads, push the lead in toward the instrument and then pull
back on the lead to fully remove lead.
DID YOU KNOW?
Always connect the ground return clip first and double check that both clips
have a solid connection to the DUT.
START
From the Perform Test screen,
select the MENU soft key twice.
On the Test Settings screen,
select the TEST soft key.
Select EDIT once you have
chosen the correct parameter.
Use the + or - soft keys to adjust
the parameter value.
GND Settings
M1-1 1.0s
____________________________
Menu
A
mÍ
M1-1 GND
Settings
________________
5.00A
120mÍ
0mÍ
1.0s
Results
Test
System
Exit
Current 5.00A
Max Lmt 120mÍ
Min Lmt 0mÍ
Dwell 1.0s
Frequency 60Hz
Offset 0mÍ
Connect OFF
Edit
Exit
>
>
Current 5.00A
Max Lmt 120mÍ
Min Lmt 0mÍ
Dwell 1.0s
Frequency 60Hz
Offset 0mÍ
Connect OFF
Edit
Exit
>
>
Current 5.00A
Max Lmt 120mÍ
Min Lmt 0mÍ
Dwell 1.0s
Frequency 60Hz
Offset 0mÍ
Connect OFF
+
-
Edit
Exit
Select a parameter to edit by
scrolling with the and
soft keys.
Current 5.00A
Max Lmt 120mÍ
Min Lmt 0mÍ
Dwell 1.0s
Frequency 60Hz
Offset 0mÍ
Connect OFF
+
-
Edit
Exit
Select ENTER to save your test
settings.
Select EXIT twice to return to the
Perform Test screen.
Current 5.00A
Max Lmt 120mÍ
Min Lmt 0mÍ
Dwell 1.0s
Frequency 60Hz
Offset 0mÍ
Connect OFF
+
-
Enter
Esc
WARNING: DO NOT TOUCH THE DEVICE UNDER TEST ONCE YOU START THE TEST.