Associated Research HYAMP III Скачать руководство пользователя страница 3

POWER

TEST

RESET

3140

GROUND BOND TESTER

SENSE+

CURRENT

RETURN

SENSE-

TEST

POWER

TEST

RESET

3130

GROUND BOND TESTER

CURRENT

INTERCONNECT

RETURN

TEST

TEST CONNECTION

CHANGE TEST SETTINGS

Setting up a test connection on the HYAMP III is straightforward, but differs 
slightly by model.

3140/3145/3160 Set Up

3130 Set Up

  

Connect the black return test lead to the black Return terminal located on  
the front panel of the instrument. 

 

 For 3140/3145/3160 models 

connect the sense lead to the Sense(-) 

terminal located just above the return terminal.

  

Clip the end of the black return test lead to the chassis ground of  
the DUT.

  

Connect the red high current test lead to the red Current terminal on the 
front panel of the instrument. 

  

For 3140/3145/3160 models

 connect the sense lead to the Sense(+) 

terminal.

  

Connect the end of the red high current test lead clip to the ground/earth 
contact of the DUT.

Push the provided current and return test leads into their mating terminal until 
seated against the rubber grommet. Avoid excessive force in seating. If the 
grommet becomes compressed, the lock feature may not activate. To release the 
lock and remove the leads, push the lead in toward the instrument and then pull 
back on the lead to fully remove lead. 

DID YOU KNOW?

 

Always connect the ground return clip first and double check that both clips 
have a solid connection to the DUT.

START

From the Perform Test screen, 
select the MENU soft key twice.

On the Test Settings screen, 
select the TEST soft key.

Select EDIT once you have 
chosen the correct parameter.

Use the + or - soft keys to adjust 
the parameter value.

GND   Settings
M1-1     1.0s

____________________________

Menu

A

M1-1   GND
  Settings

________________

    5.00A
     120mÍ
       0mÍ
     1.0s

Results

Test

System

Exit

Current   5.00A
Max Lmt    120mÍ
Min Lmt      0mÍ
Dwell      1.0s
Frequency   60Hz
Offset       0mÍ
Connect      OFF

 

 

Edit

Exit

>

>

Current   5.00A
Max Lmt    120mÍ
Min Lmt      0mÍ
Dwell      1.0s
Frequency   60Hz
Offset       0mÍ
Connect      OFF

 

 

Edit

Exit

>

>

Current   5.00A
Max Lmt    120mÍ
Min Lmt      0mÍ
Dwell      1.0s
Frequency   60Hz
Offset       0mÍ
Connect      OFF

+

-

Edit

Exit

Select a parameter to edit by 
scrolling with the       and      
soft keys.

Current   5.00A
Max Lmt    120mÍ
Min Lmt      0mÍ
Dwell      1.0s
Frequency   60Hz
Offset       0mÍ
Connect      OFF

+

-

Edit

Exit

Select ENTER to save your test
settings.

Select EXIT twice to return to the 
Perform Test screen.

Current   5.00A
Max Lmt    120mÍ
Min Lmt      0mÍ
Dwell      1.0s
Frequency   60Hz
Offset       0mÍ
Connect      OFF

+

-

Enter

Esc

WARNING: DO NOT TOUCH THE DEVICE UNDER TEST ONCE YOU START THE TEST.

4

5

Содержание HYAMP III

Страница 1: ...lips directly and always connect the return lead first You should always know when a test is being performed WARNING THISGUIDEWASCREATEDFOROPERATORSHAVINGSOMEFAMILIARITYWITHELECTRICALSAFETY TESTING AN ELECTRICAL SAFETY TESTER PRODUCES VOLTAGES AND CURRENTS THAT CAN CAUSE HARMFUL OR FATAL ELECTRIC SHOCK TO PREVENT ACCIDENTAL INJURY OR DEATH THESE SAFETY PROCEDURES MUST BE STRICTLY OBSERVED WHEN HAN...

Страница 2: ...OTE INTERFACE LINE INPUT SIGNAL INPUT CAL SIGNAL OUTPUT OPTION CAT II T6 3A 250V VOLTAGE SELECTOR INTERCONNECT RETURN CURRENT GND 115V 230V 6 3A MAX 50Hz 60Hz 115 REMOTE INTERFACE LINE INPUT 3130 3130 3130 Perform Test Screen WARNING LOCATE A SUITABLE TESTING AREA WITH A THREE PRONG GROUNDED OUTLET BE SURE THAT YOUR THREE PRONG OUTLET HAS BEEN TESTED FOR PROPER WIRING READ THE SAFETY CHECKLIST OF ...

Страница 3: ...ove the leads push the lead in toward the instrument and then pull back on the lead to fully remove lead DID YOU KNOW Always connect the ground return clip first and double check that both clips have a solid connection to the DUT START From the Perform Test screen select the MENU soft key twice On the Test Settings screen select the TEST soft key Select EDIT once you have chosen the correct parame...

Страница 4: ...nnect terminal on the HYAMP III Ground Bond tester Plug the other end into the Return terminal on the rear panel of the Hypot III Using the rear panel connections connect the following cords P N 5030DT 21 Plug the blue coded end into the Ground Bond interlock connector Plug the red coded end into Hipot signal output P N 5030DT 22 Plug the blue coded end into the Ground Bond signal output Plug the ...

Страница 5: ...DEO For additional information about these and other key features of the HYAMP III please consult the full Operation and Service Manual or call us toll free 1 800 858 TEST 8378 or 1 847 367 4077 2017 Associated Research arisafety com 1 2017 FOLLOW US ...

Отзывы: