HYAMP® Series
GROUND BOND TESTER
SENSE +
TEST
SENSE -
My Menu / Select
POWER
DATA
BARCODE
TEST CONNECTION
5
WARNING: DO NOT TOUCH THE DEVICE UNDER TEST ONCE YOU START THE TEST.
Find test leads for DUT Connections in
Section 2 of your Accessory Box.
Did You Know?
The
Data
port on the front of the instrument can be used to
download test results directly to a USB thumb drive.
The
Barcode
port is compatible with a USB barcode scanner
so you can automatically scan product barcodes and run test
sequences.
1
2
3
4
5
6
Connect the black return test lead (P/N 38490) to the black Return terminal
located on the front panel of the instrument.
Connect the black sense lead to the Sense(-) terminal located just above the
Return terminal.
Clip the end of the black return test lead to the chassis ground of
the DUT.
Connect the red high current test lead (P/N 38489) to the red Current terminal
on the front panel of the instrument.
Connect the red high current sense lead to the red Sense(+) terminal.
Connect the red clip end of the high current test lead to the ground/earth
contact of the DUT.