Page 34 of 38
Electromagnetic environment in which the ASSKEA
8.1
secretion aspirator proVisio
®
M28 may be operated
The
ASSKEA secretion aspirator proVisio
®
M28
is intended for operation in the electromagnetic
environment specified below, in which the RF disturbances are controlled. The customer or the
user of the
ASSKEA secretion aspirator proVisio
®
M28
must ensure that it is operated in such
an environment.
The environments for intended operation include professional healthcare institutions and the
home care environment. Special environments, such as the proximity of RF surgery or MRI resp.
environments in which the intensity of the EMC disturbances is high, are excluded.
Emission limits
Conducted and radiated RF emissions
CISPR 11
Harmonic distortion
IEC 61000-3-2
Voltage fluctuations/flicker
IEC 61000-3-3
Housing
Phenomenon
Test method
Immunity test levels
Electrostatic discharge
IEC 61000-4-2
±8 kV contact
±15 kV air
Radiated RF disturbances
IEC 61000-4-3
28 V/m
80 MHz to 6 GHz
80% AM at 1 kHz
Magnetic field at rated power
frequency (50 Hz)
IEC 61000-4-8
30 A/m
Power supply AC
Phenomenon
Test method
Immunity test levels
Fast transient electrical
disturbances/bursts
IEC 61000-4-4
±2 kV
100 kHz repetition frequency
Impulse voltages/surges
IEC 61000-4-5
±1 kV line-to-line
±2 kV line-to-ground
Conducted RF disturbances
IEC 61000-4-6
3V
0.15 MHz to 80 MHz
6V in ISM bands and amateur
radio bands between
0.15 MHz and 80 MHz
80% AM at 1 kHz
Voltage dips
IEC 61000-4-11
0% U
T
for ½ period
At 0°, 45°, 90°, 135°, 180°,
225°, 270°, 315°
0% U
T
for 1 period
70% U
T
for 25 periods
one-phase at 0°
Voltage interruptions
IEC 61000-4-11
0% U
T
for 250 cycles
Note:
U
T
is the AC mains voltage prior to the application of the test levels.