The device must be taken out of service and secured against inadvertently being restarted, if the presumption is that 
risk-free operation is no longer possible (see Chapter 9: Faults). 

Field Modification or other alterations to the instrument are not permitted and will void the warranty. This also applies 
to installation of spare parts. Possible conversations or alterations may only be carried out by the manufacturer. 

The operational safety of the device is only guaranteed where it is used for its intended purpose. The specification of 
the device must be adapted to the medium used in the plant. The limit values indicated in the technical data must not 
be exceeded. 

The safety information detailed in this Operating Manual, existing national regulations for accident prevention, and the 
operator’s internal regulations regarding working, operations and safety must be respected. 

The operator is responsible for all specified servicing, inspection and installation works being carried out by authorized 
and qualified specialists. 

2.4  Staff qualifications (target group assessment) 

The device may only be installed and started up by specialist staff familiar with installation, start-up and operation of 
the product. 

 

Specialist staff are defined as those individuals capable of assessing the work assigned to them on the basis of their 
specialist training, knowledge and experience of the relevant standards, and can identify possible risks.  

2.5  Signs/Safety markings 

The pressure gauge  and  packaging carry markings  detailing  the article number, measurement  range and 
manufacturer. The pressure gauge  can be provided with additional signs and safety markings advising on special 
conditions: 

  Advice on calibration 

  Safety marking pursuant to EN 837-1 

  Oil-can deleted (if used for oxygen service) 

  Silicone-free (for use in the automotive industry) 

2.6  Safety equipment 

The window of model T5500E uses multi-layer safety glass. 

This device has internal transient and reverse power protection. 

 

2.7  Environmental protection 

This device does not contain any environmental critical components.  

3  Technical data 

 

The detailed technical information can be found in the documents in the Appendix, Chapter 11. 

 

4  Labeling on the device 

The label with the serial number and type designation is located on the outside of the housing. The materials identifier 
is encoded in the type designation. 

 

Содержание t5500e

Страница 1: ...5 01EN Installation and Maintenance Instruction Manual Bourdon tube pressure gauge with integrated transmitter model T5500E for industrial application In the following configuration 100 T5500E bourdon...

Страница 2: ...g on the device 5 5 Construction and function 6 5 1 Overview 6 5 2 Description of function 6 5 3 Description of components 6 5 4 Accessories 6 6 Transport 7 6 1 Safety 7 6 2 Transport inspection 7 6 3...

Страница 3: ...ing This indicates a possibly hazardous situation where failing to follow advice may result in risks to people animals the environment and buildings Information This emphasizes key information for eff...

Страница 4: ...e will be no output 2 1 2 Freezing Prohibit freezing of media in pressure port Unit should be drained mount in vertical position to prevent possible overpressure damage from frozen media 2 1 3 Static...

Страница 5: ...t The device may only be installed and started up by specialist staff familiar with installation start up and operation of the product Specialist staff are defined as those individuals capable of asse...

Страница 6: ...ce and pointer pursuant to EN 837 1 nominal size 100 mm Erl uter ung zus tzlich ei n Hi nweis 5 3 2 Instrument connection The instrument connection is located on the bottom side of the pressure gauge...

Страница 7: ...sured the scope of the measurement range and of the protection against special conditions such as vibrations pulsations and pressure spikes A bracket must be installed to support the pressure gauge if...

Страница 8: ...nduit that contains high current AC power cables Where possible avoid running the cable near inductive equipment Units with shielded cable electrical termination connect the drain wire to the guard te...

Страница 9: ...secured against being switched on again 8 2 Check on function and recalibration Inspection of function and recalibration should be at regular intervals The precise testing cycles should be adjusted i...

Страница 10: ...acement of the pressure gauge is required in the event any of the above exist 9 4 Conduct following fault rectification See Chapter 7 3 Mounting Installation 10 Removal disposal 10 1 Safety Residue fr...

Страница 11: ...Page 11 of 11 11 2 CE Declaration of conformity...