6.3  Description of components 

6.3.1  Scale with pointer 

The pressure gauge is equipped with a dial face and pointer pursuant to EN 837-3, nominal size 100 mm or 160 mm. 

 

Erläuter ung, zusätzlich ei n Hi nweis:  

6.3.2  Instrument connection 

The instrument connection is located on the underside of the pressure gauge  and can be a threaded or flanged 

connector. 

6.3.3  Vent valve 

The vent valve for the housing is located on the top side. If the nipple is pulled out, the housing is ventilated and the 

pressure which has built up in the housing due to the influence of temperature is discharged. With the valve closed, 

protection class IP 65 is achieved. 

6.3.4  Rear wall/plug with blow-out capability 

The pressure gauge has a plug capable of blowing out on the rear wall of the housing (Model P5500) or a rear wall 

capable of blowing out (Model P6500). These act as a safety feature pursuant to EN 837-3 and simultaneously allow for 

temperature compensation for the housing, via a rubber membrane. 

6.4  Accessories 

Please contact the manufacturer regarding special tools and accessories.  

7  Transport 

7.1  Safety 

The pressure gauge  should be protected against the effects of knocks and impacts. The device should only be 

transported in the packaging provided, to protect against glass breakage. The device should only be transported in a 

clean condition (free of residues of measuring media). 

7.2  Transport inspection 

The delivery must be checked for completeness and damage during transport. In the event of damage during transport, 

the delivery must not be accepted, or only accepted subject to reservation of the scope of the damage being recorded 

and, if necessary, a complaint initiated. 

7.3  Storage 

The pressure gauge must be stored in dry, clean conditions, within a temperature range of -40 to +70 °C, protected 

against direct exposure to sunlight and protected against impact damage. 

8  Assembly/Installation 

8.1  Safety 

To ensure safe working during installation and servicing, suitable shut-off valves must be installed in the plant (see 6.4 

Accessories), enabling the device: 

 

To be depressurized or taken out of operation; 

 

To be disconnected from the mains supply for repair or checks within the relevant plant; 

 

Or to enable function tests of the device to be performed “on site”. 

During the works to mount/install the gauge, the plant must be protected against being switched back on. 

8.2  Preparations (requirements for the installation location) 

 

A check on suitability of the device for the medium to be measured, the scope of the measurement range and of the 

protection against special conditions such as vibrations, pulsations and pressure spikes. 

 

A bracket must be installed to support the pressure gauge if the metering pipe is not able to provide adequate 

support. 

 

Devices with a blow-out require a minimum spacing to the rear (20 mm). 

Содержание P5500

Страница 1: ...tion Manual Diaphragm pressure gauge model P5500 and P6500 for explosion risk areas pursuant to Directive 2014 34 EU ATEX In the following configuration P5500 P6500 ATEX diaphragm pressure gauge without switching contact P5500 P6500 I ATEX diaphragm pressure gauge with inductive proximity switches ...

Страница 2: ...tching contact 6 3 2 P5500 P6500 I with inductive proximity switches SJ2 N 6 4 Technical data 7 5 Labeling on the device 8 5 1 Labeling on the device for explosion risk areas ATEX 8 6 Construction and function 8 6 1 Overview 8 6 2 Description of function 8 6 3 Description of components 9 6 4 Accessories 9 7 Transport 9 7 1 Safety 9 7 2 Transport inspection 9 7 3 Storage 9 8 Assembly Installation 9...

Страница 3: ...12 11 2 Removal 12 11 3 Disposal 12 12 Appendix 13 12 1 Declaration of conformity model P5500 13 12 2 Declaration of conformity model P5500 with inductive switches 15 12 3 EU design type test certification for inductive proximity switches of types SJ2 N 17 12 4 Data sheet for diaphragm pressure gauge P5500 P6500 28 ...

Страница 4: ...ded in such an eventuality 1 4 Copyright This Operating Manual may only be copied and passed on as a complete document without the special permission of the publisher 1 5 Warranty For the product described here we offer a warranty pursuant to Section 6 Guarantee in Respect of Defects in our General Terms and Conditions of Delivery and Payment 1 6 Manufacturer s address customer services Ashcroft I...

Страница 5: ...inspection and installation works being carried out by authorized and qualified specialists The operator is responsible that the device is used in the correct ATEX zone 2 4 Staff qualifications target group assessment The device may only be installed and started up by specialist staff who are qualified for installation start up and operation of the product Specialist staff are people who are able ...

Страница 6: ...r a change in differential pressure between 10 and 90 of the measuring range and a pulse frequency of 0 06 Hz the temperature increase is 10 K To avoid additional temperature increase the devices should not be exposed to direct exposure to sunlight when in operation For the non electrical part of the devices the standards EN 60079 0 EN 60079 31 EN ISO 80079 36 and EN ISO 80079 37 are applicable wi...

Страница 7: ... data per contact max internal capacitance Ci max 30 nF max internal inductance Li max 100 μH The permitted limit values for Ui Ii and Pi for the intrinsically safe feed power circuits depend on the proximity switch type They are to be found in the respective EU design type test certifications For the non electrical part of the devices the standards EN 60079 0 EN 60079 31 EN ISO 80079 36 and EN IS...

Страница 8: ...e proximity switch P5500 I ATEX 6 Construction and function 6 1 Overview 1 Diaphragm 2 Rod 3 Movement 6 2 Description of function The sensing element the concentrically rippled diaphragm element is clamped in place between two flanges and charged with the medium on one side Under the pressure acting on it there is an elastic displacement out of the normal position This displacement is proportional...

Страница 9: ...y be transported in a clean condition free of residues of measuring media 7 2 Transport inspection The delivery must be checked for completeness and damage during transport In the event of damage during transport the delivery must not be accepted or only accepted subject to reservation of the scope of the damage being recorded and if necessary a complaint initiated 7 3 Storage The pressure gauge m...

Страница 10: ...ter is used as the measurement medium the device must be frost protected Safety notice Only mount using the correct open jawed wrench and do not twist the device itself 8 3 2 Electrical connection Take note of the electrical data in the EU design type test certification and the locally applicable regulations and guidelines for installing and operating electrical plant in explosion risk areas e g E...

Страница 11: ... Open contacts with the indication rising in a clockwise direction Contact assignment 1st contact left target value indicator 2nd contact middle target value indicator 3rd contact right target value indicator 8 5 Subsequent relocation of the gauge by the customer Recommendation Do not remove the pressure gauge from one metering point and fit it in a different place as there is a risk of the measur...

Страница 12: ...nt or loose pointer Cracked window Leaks when the device is filled Damage to housing Indications that the measurement system seal is imperfect discoloration to dial display or of filling liquid In these instances replacement of the pressure gauge is always required 10 4 Conduct following fault rectification See Chapter 8 3 Mounting Installation 11 Removal disposal 11 1 Safety Residues of measuring...

Страница 13: ...Page 13 of 28 12 Appendix 12 1 Declaration of conformity model P5500 ...

Страница 14: ...Page 14 of 28 ...

Страница 15: ...Page 15 of 28 12 2 Declaration of conformity model P5500 with inductive switches ...

Страница 16: ...Page 16 of 28 ...

Страница 17: ...Page 17 of 28 12 3 EU design type test certification for inductive proximity switches of types SJ2 N ...

Страница 18: ...Page 18 of 28 ...

Страница 19: ...Page 19 of 28 ...

Страница 20: ...Page 20 of 28 ...

Страница 21: ...Page 21 of 28 ...

Страница 22: ...Page 22 of 28 ...

Страница 23: ...Page 23 of 28 ...

Страница 24: ...Page 24 of 28 ...

Страница 25: ...Page 25 of 28 ...

Страница 26: ...Page 26 of 28 ...

Страница 27: ...Page 27 of 28 ...

Страница 28: ...site see 1 6 Manufacturer s address customer services This Table refers to specific documents Model Description Document P5500 Diaphragm type pressure gauge model P5500 G1 P5500 P6500 Solid front diaphragm type pressure gauge model P6500 G1 P5500 K5500 Electrical contact devices for pressure and temperature gauges G1 K5500 ...

Отзывы: