background image

of 20

11

the pressure. To check operation pull the shuttle valve 

outward, close the release valve and operate the pump 

handle several times. When handle resistance is felt, push 

the shuttle valve inward and continue pumping within the 

range of the test device.
Entrapped air will prevent the pump from operating in the 

high pressure valve position or cause it to achieve only 

partial pressure. Repeat the above bleeding procedure 

as necessary to assure all air is removed. Once air is re-

moved the pump will continue to operate without further 

attention providing the reservoir level is maintained.
3.5.3 Making The Test 

Turn displacement piston handle (41) to mid-position, so 

that it may be used for setting an exact pressure on the 

gauge. The displacement piston has internal stops which 

prevent unscrewing or accidental loss of pressure.
Clockwise rotation of displacement piston will produce 

an increase in pressure; counterclockwise will decrease 

 pressure.
Prior to taking readings, both gauges should be fin-

ger-tapped lightly at the center of the gauge face, to elim-

inate any movement friction. Note the pressure readings 

on the test gauge and the gauge under test. If the pres-

sure indicated on the gauge under test is not equal (within 

the tolerance of the gauge) to the pressures of the master 

test gauge, the gauge being tested requires calibration.
CAUTION: 

Do not pump handle to pressures greater than the pres-

sure range of the gauges connected to the test pump, as 

this may damage the gauges.
3.5.4 To Release Pressure 

Open release valve slowly (37), until pressure returns to 

zero. Do not loosen any connections until pressure in the 

gauge tester has reached zero, as indicated on the test 

gauge.
If additional gauges are to be tested, close the release 

valve when the pressure reaches zero. This will prevent 

complete drainage of oil in the tester back to the reservoir.
Unseal the gauge that has been tested by unscrew-

ing adapter nut (62) from connector (60) several turns. 

Unscrew the gauge from bushing (59) if used, or connec-

tor (60), and remove.
Remove the test gauge in the same manner as the gauge 

under test.
3.5.5 To Dismantle 

To replace test pump in case, remove gauges and both 

tube assemblies. Put shipping plugs into pump and tighten. 

Close filler plug (35).
3.5.6 Operating Fluids 

Standard Test Pump: 

Any medium weight oil may be used (including automotive 

oils S.A.E 10, 20 or 30) S.A.E. 20-W recommended.
Hydraulic Test Pump: 

Any hydraulic fluid (silicate or phosphate base). Skydrol or 

Pydraul is suitable.
Oxygen Test Pump: 

Distilled or demineralized water with a compatible corro-

sion inhibitor added. A .1% solution of sodium dichromate 

is a suitable inhibitor.
For normal operation it is not necessary to change the 

weight of oil for various pressures. A lighter oil may be 

used where low pressures are being checked, a heavier 

oil where higher pressures are being tested.

The reservoir can be refilled while the test pump is operat-

ing under pressure.
3.5.7 Caution 

Standard testers designed for oil service may not be used 

with water for oxygen service.
3.5.8  

Hydraulic service testers should not be filled with water or 

any oil other than hydraulic fluid. Serious pump failure may 

occur due to O-ring damage.
4.0   FACTORS AFFECTING OPERATION OF  

DEADWEIGHT TESTERS

4.1  Deadweight Tester Accuracy 

Excellent accuracy is possible using the deadweight tes-

ter through analysis and control of certain factors. If the 

tester’s rated accuracy of 1/10th of 1% is adequate, then 

the nominal pressure (sum of the denominations of the 

weights loading the piston) may be assumed to be correct. 

The pressure normally developed is determined by this 

formula: Deadweight Tester Pressure =
 

  Mass of weights plus piston mass

 

  Effective area of piston and cylinder

 

    P = (M-ph)  g

 

               A     gs 

= pressure 

M  = mass of the load on the piston 

= effective area of the piston in sq. inches, or sq. cm 

g/gs = ratio of the value (g) of gravity at the point of use to  

 

  the standard value of gravity (gs) 

= density of liquid used in test 

= difference in level between gauge being tested, and  

 

  the bottom of the piston

The deadweight tester is capable of measuring pressures 

to proper corrections after analyzing these factors.
4.2  Gravity, Calibrated Weights 

Weights furnished are calibrated at standard gravity of 

980.665 gals. If precise accuracy is required, the error 

introduced by change of weight due to change in gravity at 

the locality should be calculated and included in results.
The mechanism of an Ashcroft Pressure Gauge includes 

a Bourdon tube and geared movement, which is unaf-

fected by variations in gravity. Conversely, the pressure de-

veloped by a deadweight tester is proportional to the value 

of gravity. Readings of this type pressure gauge will cor-

respond to those of a deadweight tester when the tester 

is subjected to standard gravity (gs = 980.665 gals. in the 

International System). In southern sections of the United 

States, the value of gravity may be several thousandths 

less than the standard value, if the latitude (Ø) and the 

elevation above sea level (a) for an area are known, the 

approximate value of (g) in gals. may be calculated from 

this formula:
 

g = 980.632-2.586 COS 2Ø 

 

      +.003 COS 4Ø - .000094a

Ø = Latitude (Degrees) 

a  = Elevation above sea level (ft)
4.3  Effective Measured Area: Chamber 

The effective area of the deadweight tester may be de-

termined by the average of the cross sectional area of 

the piston and the area of the cylinder bore. This effective 

area is affected somewhat by temperature, and by the 

elastic distortion of the piston and cylinder when pressure 

is being applied. The effective area of a stainless steel 

piston and cylinder increases approximately .068% with 

Содержание 1305D

Страница 1: ... Manual for the ASHCROFT Type 1305D Deadweight Tester andType 1327D Portable Pump Installation and Maintenance Manual for the ASHCROFT Type 1305D Deadweight Tester andType 1327D Portable Pump I M002 10095_RevF_ 250 1526 _09 20 22 ...

Страница 2: ...of 20 2 ...

Страница 3: ...0018 New York NY 10017 1 Pressure Select a range so that the maximum ap plied pressure will never exceed the upper range limit 2 Vibration Excessive vibration could cause a loosen ing of components and abnormal wear resulting in loss of instrument accuracy or failure to provide valid data 3 Temperature Operation of the instrument in an en vironment where temperatures are in excess of design rating...

Страница 4: ...nt mass being acted upon by gravity For this rea son the masses supplied with the tester are stamped with two pressure values the value being contingent on the effective area of the piston and cylinder assembly selected 2 1 3 Specifications 1305D Accuracy Combined tolerance of weights and piston and cylinder assemblies within 0 1 of reading Weight toler ance within 0 05 of mass Piston and cylinder...

Страница 5: ...1 2 1 26 175 80 Model Piston Assembly Piston Net Number Pressure Range Piston Area Value Number of Weights by Value Weight L 5 L 1 L 2 L 4 L 5 bar Type Low High Low High Low High H 2 5 H 5 H 10 H 20 H 25 lbs kg 1305DB 20 1305DBH 20 1 20 5 2 5 1 1 2 1 2 63 29 1305DB 100 1305DBH 100 1 20 5 100 5 2 5 1 1 2 1 2 65 30 1305DB 150 0613 in 2 0123 in 2 1305DBH 150 1 30 5 150 3954 0791 5 2 5 1 1 2 1 4 75 34...

Страница 6: ...andard 1 4 inch internal NPT and 1 2 inch internal NPT fitting adapters are supplied Operating Fluid 1327D Light grade machine oils au tomotive petroleum base SAE 20 oils or other equivalent fluids suitable for use with Buna N O ring materials 1 5 pints required 7 liters 1327DH Most hydraulic oils of phosphate ester base brake fluids skydrol pydraul etc suitable for use with Butyl or Ethylene Prop...

Страница 7: ...DMG 250 1327DMGH 250 X X 0 10 0 40 0 250 40 18 1327DMGO 250 1327DMG 600 1327DMGH 600 X X 0 10 0 40 0 250 0 600 42 19 1327DMGO 600 1327DBG 10 1327DBGH 10 X X 0 10 36 16 1327DBGO 10 1327DBG 40 1327DBGH 40 X X 0 10 0 40 38 17 1327DBGO 40 bar 1327DBG 250 1327DBGH 250 X X 0 10 0 40 0 250 40 18 1327DBGO 250 1327DBG 600 1327DBGH 600 X X 0 10 0 40 0 250 0 600 42 19 1327DBGO 600 1327DAG 1000 1327DAGH 1000 ...

Страница 8: ...ndle mo tion Repeat this action until no air bubbles are observed in the fluid flow Note the bleed screw must be closed when the handle is being raised Push the shuttle valve in towards the pump body and continue pumping to the desired test pressure Open the release valve 37 to vent the pressure To check operation pull the shuttle valve outward close the release valve and operate the pump handle s...

Страница 9: ...of 20 9 Figure 2 1 Ashcroft Portable Deadweight Tester 1305D ...

Страница 10: ...he gauge to the desired position and re tightening nut adapter Repeat the above procedure to connect the gauge being tested to the offset pipe assembly 64 or 66 3 4 3 Bench Space Required Allow a space 10 inch 2 5 cm width 23 inch 5 8 cm depth 22 inch 5 6 cm height allows for testing gauges up to 8 5 inch 250 mm size 3 4 4 Storage Space Required 101 2 inch 27 cm width 221 2 in 57 cm depth 10 inch ...

Страница 11: ...ght of oil for various pressures A lighter oil may be used where low pressures are being checked a heavier oil where higher pressures are being tested The reservoir can be refilled while the test pump is operat ing under pressure 3 5 7 Caution Standard testers designed for oil service may not be used with water for oxygen service 3 5 8 Hydraulic service testers should not be filled with water or a...

Страница 12: ...uracy of the finished weights better than 1 20th of 1 To maintain deadweight tester accuracy handle the weights with care and keep the piston and cyl inder in clean condition The tester should be flushed with a solvent occasionally preferably every six months so that operating fluid is al ways clean This will prolong component life and provide protection against possible sticky action between the ...

Страница 13: ...f the tester body assembly casting When replacing O ring packings coat them with suitable lubricant to pre vent sticking and tearing during assembly and tightening of connection 5 14To Reassemble Body assembly on drip pans Replace body assembly 2 on drip pan 3 using bolts 25 5 15 Replacing Displacement Valve a Place O ring packing between back up washers b Attach back up washers and O ring packing...

Страница 14: ...of 20 14 Figure 2 2 Ashcroft Portable Test Pump Type 1327D ...

Страница 15: ...ylinder 3 Damaged piston plate damaged cylinder threads or damage to any component part that results in inferior performance or malfunctioning of the unit The piston cylinder assembly is available as a unit only This assures the user of maximum accuracy in his Ashcroft Deadweight Tester and maintains certified trace ability to the N I S T 6 0 WARRANTY SHIPPING INSTRUCTIONS HOWTO ORDER 6 1 Warranty...

Страница 16: ...ate with deadweights drops rapidly Unable to hold constant pressure Worn piston and cylinder assembly Use a heavy oil temporarily Replace piston cylinder Stacked dead weights wobble when spinning piston plate Damaged dead weights Check dead weights for visible damage bends dents nicks etc and alignment Piston plate as sembly will not spin Too heavy an oil grade being used Flush fill unit with prop...

Страница 17: ...of 20 17 4A ...

Страница 18: ...d Valve 27 70 52 1 LEE236 Nameplate 4 70 58 1 LK227 Pipe Extension Assembly 180 00 Qty Unit Item Per List No Asmy Part No DescriptionPrice 59 2 LD227A Bushing2 9 50 60 2 LN141 Adapter 25 40 61 4 LD186 Collar 9 50 62 4 LD117 Nut Adapter 25 20 63 1 LH227A Extension Pipe 67 00 64 1 LJ227 Lower Connection Pipe Assembly 65 1 LF227A Lower Connection Pipe 104 00 66 1 LG227A Back Connection Pipe 104 00 67...

Страница 19: ... psig 100 500 PSI WEIGHT SETS Weight Set Number of Weights with by High Pressure Value Part Carrying 25 50 100 200 500 Number Case psi psi psi psi psi LBH199C 1000 psi 1 3 2 3 LBH199B 2000 psi 1 3 2 3 2 LBH199A 3000 psi 1 3 2 3 4 LBH199 5000 psi 1 3 2 3 8 LBH199D 5000 to 10 10000 psi For complete 10 000 psi weight set order both part no LBH199 and LBH199D AVAILABLE WEIGHTS WEIGHT SETS ...

Страница 20: ...change without notice All sales subject to standard terms and conditions Ashcroft Inc I M002 10095_RevF_ 250 1526 _09 20 22 Ashcroft Inc 250 East Main Street Stratford CT 06614 5145 U S A Tel 203 378 8281 email info ashcroft com www ashcroft com ...

Отзывы: