Syringe Tip Sterilization:

1) Remove contaminated syringe tip.
2) Remove all visible signs of contamination

before autoclaving.

3) Autoclave tip at 132° C (270° F) for ten

minutes.

4) Sterilize between each patient use.

NOTE:

Since only the tips can be autoclaved, it is

recommended that the air/water syringe be bagged
with a disposable, single-use plastic sleeve
between each patient use.

AA-21SX Air Vacuum System

Handpiece Control-

position

-

‘ON’

When the unit is connected to a

compressed air source, and the vacuum bottle assembly is attached to the
air receptacle on the side of the unit case, the Air Vacuum System should be
fully functional. Insert a standard disposable vacuum tip into the hose end.
Depress button on hose end to activate vacuum. Observe waste bottle while
in use and dispose of waste when bottle is approximately 3/4 full.

Avoid

 

waste

 

flow-back

 

during

 

use

 

by

 

keeping

 

the

 

ON/OFF

 

button

 

depressed

 

prior

 

to

 

entering

 

the

 

oral

 

cavity

 

and

 

until

 

the

 

hose

 

end

 

has

 

been

 

removed

 

from

 

the

 

patients

 

mouth.

 

Clean

 

the

 

entire

 

hose

 

assembly

 

and

 

waste

 

bottle

 

after

 

each

 

use

 

by

 

evacuating

 

a

 

cleaner/disinfectant

 

through

 

the

 

system

 

and

 

thoroughly

 

scrubbing

 

entire

 

assembly.

 

CAUTION

: Use only NON-foaming 

cleansers in the vaccum lines.

The ADU-08SXE unit includes one manual hand-

piece control with water coolant and pressure gauge. Attach any 4-hole
Midwest style highspeed, lowspeed, or air sonic scaler to the control tubing
and place in the holder on the front of the unit. Depress the disc foot control
for handpiece operation. For water coolant, switch the ‘Water Coolant’ toggle
to the

. Water coolant will activate with handpiece operation.

ON/OFF

BUTTON

INSERT DISPOSABLE

VACUUM TIP

ADJUSTMENTS:

Master Pressure-

Water Pressure-

Syringe Flow-

Handpiece Drive Air-

The master pressure regulator controls the incoming

supply pressure. It is located behind the vacuum holder, underneath the
control panel, inside the unit. This regulator is factory set and should not
require further adjustment. If further adjustment is needed, turn knob
clockwise to increase pressure and counter clockwise to decrease.

The water pressure regulator controls the pressure inside

the water supply bottle. It is located to the left side of the water bottle, under-
neath the control panel, inside the unit.

The flow block controls the air flow to the three-way air/water

syringe. It is located directly behind the syringe, underneath the control
panel, inside the unit. Using a small screwdriver blade, turn the right
adjusting screw clockwise to decrease air flow and counterclockwise to
increase.

If further adjustment is needed, turn

knob clockwise to increase pressure and counter clockwise to decrease.

The flow block controls the drive air to the handpiece.

It is located directly behind the syringe, underneath the control panel, inside
the unit. Using a small screwdriver blade, turn the left adjusting screw
clockwise to decrease drive air pressure, and counterclockwise to increase.
The pressure gauge on the front panel indicates handpiece running pressure.

The water coolant flow knob controls the water coolant

flow to the handpiece control. It is located directly behind the handpiece
holder, underneath the control panel, inside the unit. Turn the knob clock-
wise to decrease water coolant flow, and counterclockwise to increase.

Water Coolant Flow-

ADU-08SXE CONTROL PANEL

HighSpeed
LowSpeed
Air Scaler

Attach

Handpiece

Depress

Foot Control

WATER

BOTTLE

MOUNT

MASTER

REGULATOR

WATER

REGULATOR

WATER

COOLANT

AIR ADJUST

SEALANT

Express

ON - OFF

FLOW

SYRINGE
AIR

H.P. DRIVE

AIR

ADU-08SXE

HANDPIECE PRESSURE

MASTER

PRESSURE

ADJUSTMENT

SYRINGE AIR

ADJUSTMENT

HANDPIECE DRIVE

AIR ADJUSTMENT

WATER

PRESSURE

ADJUSTMENT

H.P. WATER

COOLANT

FLOW ADJUST

UNIT OPERATION:

Depress the right button for air operation,

and the left button for water operation. Depressing both buttons will create a
mist. The syringe features quick-change autoclavable tips: To remove a tip,
press on the locking collar surrounding the tip socket and pull the used tip
straight out of the socket. To insert a new tip, press locking collar and push tip
into socket as far as it will go. Release collar and gently tug on tip before
using to ensure that tip is securely locked into socket.

TA-90D Air/Water Syringe

-

SYRINGE TIP REMOVAL

SYRINGE TIP REMOVAL

DEPRESS

LOCKING COLLAR

DEPRESS

LOCKING COLLAR