background image

Page 5

WARNING:

Do not turn the halogen Lamp ‘On’

without the Guard/Lens Assembly installed. The

Lamp emits harmful UV radiation and damage to

eyes and skin could result if Assembly Lens (UV

Filter) is not in place. 

WARNING:

Halogen bulb is HOT ! Operating the

Lamp without the Guard installed can cause burns!

Step 5 - Cover Lamp:

1.

Retract (turn CCW) the Heat Shield Tube away from

the Rear Casing to enclose the Lamp. Back the Tube

out until its forward lip protrudes approx. 0.075 inch

beyond the front edge of the Lamp. 

IMPORTANT:

Not retracting the Shield Tube forward may cause the

Lamp to come in contact with the glass Lens on the

Guard Assembly. Ensure that the Lamp reflector is

centered within the Shield and does not contact any

part of the Shield Tube or glass Lens on the Guard. 

Step 6 - Replace Guard Assembly:

1.

Grasp the Guard Assembly as indicated and carefully

match screw threads with those on the Rear Casing.

IMPORTANT:

Do not touch or apply pressure on the

front surface of the Guard Assembly Lens. Fingerprints

left on the Lens will attenuate the light beam. Do not

remove the two retaining clips and screws which attach

the Lens and O-ring to the Guard.
Install (turn CW) Guard Assembly onto the Casing. If

the Guard does not easily screw onto the Casing, the

threads are likely misaligned - Do Not Force! Back the

Guard off and try again. 

IMPORTANT:

When the

Guard Assembly Lens comes in contact with the Heat

Shield Tube, stop tightening! Damage to Lens and

other internal components may occur if overtightened.

Step 5 - Cover Lamp

Step 6 - Replace Guard

CCW

Problem:
Power is 'ON', but lamp does not

light:

Light intensity appears dimmer

than normal:

Blown fuse:

Correction:
-Replace Fuse: 5 x 20mm 250 mA (for 110V), or 5x20mm 160 mA (for 220V)

-Replace Lamp.

-Check plug and electrical source.

-Turn intensity up.

-Check that voltage switch matches power source

-Clean Lens (Note: Before cleaning, turn Unit ‘Off’ and allow to cool. Do not

remove the retaining clips and screws which attach the Lens to the Guard)

-Check that Lamp pins are straight and are not bent. (Bent pins can cause a

short circuit and blow the fuse if they make contact with the Heat Shield Disk.)

TROUBLESHOOTING:

HEAT SHIELD TUBE

GUARD

ASSEMBLY

0.05”

TO

0.10”

HEAT

SHIELD

LAMP

CHANGING THE FUSE 

(Fig. 2)

1.

IMPORTANT: Turn ALU-29CF Power Switch ‘Off’

and unplug unit from power outlet before changing

the fuse.

Use a slot-head screwdriver to pry the

Fuseholder out of the Power Inlet Connector, located

on bottom of the electrical housing. Install new fuse

(see type below) and replace holder in Connector,

applying pressure on holder until it firmly snaps into

position. Set Dual Voltage Switch to proper voltage.
110VAC FUSE: 5x20mm 250 mA, slo-blo type

220/250VAC FUSE: 5x20mm 160 mA, slo-blo type

Figure 2 - Fuse Replacement

FUSEHOLDER

DUAL VOLTAGE SWITCH

HEAT SHIELD

DISK

Содержание COMMAND-AIR ALU-29CF

Страница 1: ...ALU 29CF Portable Light NSN 6520 01 446 4170 OPERATION MAINTENANCE MANUAL PARTS LIST COMMANDlAIR ...

Страница 2: ...eristics 3 Items Furnished 3 Operation Instructions 3 Set Up 3 Lamp Controls 4 Maintenance 4 Lamp Replacement 4 Changing the Fuse 5 Troubleshooting 4 Parts Diagram 6 Parts List 7 TABLE OF ILLUSTRATIONS Figure 1 Setup 3 Figure 2 Fuse Replacement 5 Figure 3 Parts Diagram 6 ...

Страница 3: ...nt SETUP Fig 1 1 Remove the ALU 29CF Light from its packing and carrying case Unfold the tripod legs to a horizontal position slide to the bottom of the center post and secure in position by firmly tightening the thumb screw Set light on a stable even floor surface 2 Loosen knob on horizontal boom and position where desired Tighten to set Rotate handle counterclockwise on telescoping center column...

Страница 4: ...lign with the socket receptacles which are accessible through the opening in the Heat Shield Disk Carefully press the pins straight into the receptacles until the base of the Lamp is firmly seated against the Disk If the Lamp pins are properly aligned the Lamp will seat easily CAUTION Do not press hard on Lamp it could break and cause injury NOTE The pin receptacles are offset at a slight angle Do...

Страница 5: ... Force Back the Guard off and try again IMPORTANT When the Guard Assembly Lens comes in contact with the Heat Shield Tube stop tightening Damage to Lens and other internal components may occur if overtightened Step 5 Cover Lamp Step 6 Replace Guard CCW Problem Power is ON but lamp does not light Light intensity appears dimmer than normal Blown fuse Correction Replace Fuse 5 x 20mm 250 mA for 110V ...

Страница 6: ...Page 6 Aseptico ALU 29CF Command Air Portable Light Figure 3 Parts Diagram PCB Assembly 16 12 26 28 10 11 15 17 13 24 23 25 30 29 1 2 5 6 19 21 14 18 8 9 3 7 4 22 27 31 33 34 20 21 22 28 32 ...

Страница 7: ...Knob 1 4 Blk w Gray Insert 850032 1 Pot 100K Linear 810123 1 Switch Rocker SPST 830002 1 Voltage Selector Switch 840042 1 Power Inlet Fused Snap In 840035 1 Fuse 5x20mm 250 mA for 110V 830049 2 Fuse 5x20mm 160 mA for 220V 830046 1 Capacitor 100V 0 22MFD 810108 1 Transistor R6004LT Triac 880037 1 Resistor 1 4 w 5 1 0K 810006 1 PCB ALU 29CF 550043 1 Transformer 115 230V 13 6V 2A 800100 1 Chassis Bas...

Страница 8: ...s sole option and discretion to repair or replace any defective component or product in part or whole Aseptico shall be the sole arbiter of such action In the event of alleged defect under warranty the purchaser is to notify Aseptico s Customer Service Department promptly Customer Service will provide instructions usually directing that the product be returned for service Shipment to Aseptico and ...

Отзывы: